General Monitor - GE Corometrics 250cx Series Service Manual

Hide thumbs Also See for Corometrics 250cx Series:
Table of Contents

Advertisement

General Monitor

Category
Power Requirements
Nominal Line Voltage:
Line Frequency:
Power Consumption (maximum):
Chassis Leakage:
Physical Characteristics
Height:
Width:
Depth:
Weight:
Environmental Conditions
Monitor:
Ambient Temperature:
Relative Humidity:
Atmospheric Pressure:
1
Strip Chart Paper
:
Ambient Temperature:
Relative Humidity:
Atmospheric Pressure:
Certification
ANSI/AAMI EC13-1992:
UL-2601.1:
CUL:
1
Paper operating environmental conditions are for a period of less than one month. Paper storage environmental conditions are for extended storage.
Revision C
Technical Specifications General Monitor
Table 1. General Monitor Technical Specifications
100VAC
120 VAC
50/60 Hz
50/60 Hz
100 W
100 W
<300 µA
6.7 in (17.0 cm)
16.7 in (42.4 cm)
17.5 in (44.4 cm)
22.0 lbs (10.9 kg) approx.
Operating
50°F to 104°F (10°C to 40°C)
10% to 95%, non-condensing
700–1060 mbar (525–795 mmHg)
50°F to 104°F (10°C to 40°C)
30% to 70%, non-condensing
700–1060 mbar (525–795 mmHg)
Complies with all areas except those listed below:
3.1.2.1e: Heart Rate Meter Accuracy and Response to Irregular Rhythm (not tested)
3.2.6.1: Range of QRS wave amplitude and duration
3.2.7: Range and accuracy of heart rate meter (4.2.7 f: input rate of 300 bpm.)
3.2.8.1: Lower Alarm Limit (The lowest alarm limit on the 250cx Series is 35 bpm.)
3.2.9.7a: Output Display a) Channel Width
3.2.9.8c: Impulse Response
3.2.9.12: Pacemaker Pulse Display capability
Classified to UL-2601.1
Medical electrical equipment classified by Underwriter's Laboratories, Inc., with respect to
fire, shock, and mechanical hazards in accordance with UL-2601.1.
Classified with respect to electric shock, fire, mechanical, and other specified hazards
only, in accordance with CAN/CSA C22.2 No. 601.1
250cx Series Maternal/Fetal Monitor
2036947-001
Technical Specifications
220 VAC
230 VAC
50/60 Hz
50/60 Hz
0.4 A
0.4 A
Storage
14°F to 131°F (–10°C to 55°C)
0% to 95%, non-condensing
700–1060 mbar (525–795 mmHg)
< 80°F (< 26.5°C)
45% to 65%, non-condensing
700–1060 mbar (525–795 mmHg)
240 VAC
50/60 Hz
0.4 A
A-3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents