Installation: Service Mode Screens
This field is used to set the language shown on the display and printed on the strip chart
recorder paper. Use the Trim Knob control to cycle through the available options:
English, Spanish, French, German, Dutch, Swedish, Italian, Danish, Finnish, and Norwegian.
(The monitor is factory set for the country in which it is being used.)
This field is used to define the line power frequency for the country in which the
monitor is being used. Use the Trim Knob control to alternate between 50 Hz and 60
Hz. (The monitor is factory set for the country in which it is being used.)
This field is used to set the heart rate scale for the strip chart paper. Use the Trim
Knob control to alternate between 30–240 bpm/cm and 50–210 bpm/cm. (The monitor
is factory set for the country in which it is being used.) The MSpO
is also affected by this setting. When the heart rate scale is 30–240, the MSpO
expanded scale is 60–100% (10%/cm). When the heart rate scale is 50–210, the
MSpO
expanded scale is 50–100% (12.5%/cm).
2
The monitor offers a choice of font sizes to print annotations.
Small: increases print speed.
Medium: a compromise between the large and small sizes.
Large: fosters readability.
The monitor provides an option for the printing of blood pressure and MSpO
on the strip chart paper:
Real-Time: Values are printed according to the actual clock time (9:33, 9:48,
10:03, etc.).
Chart-Style: Values are printed on standard clock quarter (9:00. 9:15, 9:30:, 9:45,
etc), half (9:00, 9:30, 10:00, 10:30, etc), and whole hour marks (9:00, 10:00,
11:00, 12:00, etc).
This field enables/disables an audible tone to indicate a low-paper or out-of-paper
condition.
Off: no audio tone sounds with either condition; however, the visual indicator is
still present (the recorder LED blinks).
Low/Out: sounds a tone when either LOW BATTERY or OUT OF PAPER condition
exists.
Out Only: sounds a tone only when the OUT OF PAPER condition exists.
250cx Series Maternal/Fetal Monitor
2036947-001