Speci Cations - 3M Bair Hugger 675 Service Manual

Warming unit
Hide thumbs Also See for Bair Hugger 675:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Acronym Definitions
EMI
Electromagnetic Interference
EMC
Electromagnetic Compatibility
ESD
Electrostatic Discharge
IV
Intravenous
LED
Light-emitting Diode
LCD
Liquid Crystal Display
RF
Radio Frequency
PCS
Proximal Control Sensor
PMS
Proximal Monitor Sensor
RPM
Revolutions Per Minute
Physical Characteristics
Dimensions
14.5" high x 10.5" deep x 11" wide
(36.2 cm high x 26.7 cm deep x 28 cm wide)
Weight
9.93 lb (4.5 kg)
Relative noise level
50 dBA
Hose
Flexible
Length: 80" (203 cm)
Weight: 1.51 lb (0.686 kg)
Diameter: 2" (5.08 cm)
Unit Intake Filter
MERV 14*
Recommended Filter Change
Every 12 months or 500 hours of use.
*See "Frequently Asked Questions on Filtration" section
Recommended operating environment
Temperature
15°C-25°C (59°-77°F)
Humidity
Up to 80% relative humidity (non-condensing)
Altitude
Max 2,000 m
Operating temperatures
Temperature control
Electronically controlled
Heat generated
2217 BTU/hr (average), 650 W (average)
Time to reach operating
2 - 5 minutes (dependent on blanket model)
temperature
Time required for the contact surface temperature to
heat up from 23 ± 2°C to 37°C (73 ± 2°F to 99°F)
Storage/transport
-20°C to 45°C (-4°F to 113°F)
temperature
Store all components in a cool, dry place when not in use.
Safety System
Thermostat
Independent electronic circuit; thermal cutoff shuts the heater
OFF to ensure hose end air remains below 56°C (53°C ±
3°C typical).
Alarm system
Over-temperature (≤56°C, 53°C ± 3°C typical): red Over-Temp
indicator symbol will flash alternately with the "Wrench"
symbol, alarm sounds, heater and blower shut down, operating
indicator lights turn OFF, and control panel becomes
unresponsive. Remove from service.
Under-temperature (<3°C of selected temperature/setting
for 3 minutes): blue Under-Temp indicator symbol will flash
alternately with the "Wrench" symbol, alarm sounds, heater
and blower shut down, operating indicator lights turn OFF, and
control panel becomes unresponsive. Remove from service.
Fault
"Wrench" symbol displays and alarm sounds. Heater and blower
shut down, operating indicator light turns OFF and control
panel becomes unresponsive. Service required.
Overcurrent
Dual input fused lines. Fuses are to be accessed and/or
protection
replaced only by service personnel.
12
Alarm/Fault Codes
Alarm conditions
General fault
Over-temperature
Under-temperature
Fault code indicator
Yes on LCD screen
Filter change indicator
Yes: Triggered by 500 hours of use
Electrical Characteristics
Heating element
1190 W Resistive
Heater control
3 stage heater
flicker free
Temperature settings
Ambient
32° C (89.6° F)
38° C (100.4° F)
43° C (109.4° F)
Temperature accuracy
+/- 3° C
Temperature display
0.1° C
resolution
Leakage current
Meets UL 60601-1 and IEC 60601-1 requirements
Blower motor
Non-adjustable, one speed
Airflow: up to 44 cfm or 20.77 L/s
Operating speed: 4,700 RPM
Power consumption
Peak: 1255 W
Average: 650 W
Power cord
15 ft., SJT, 3 cond., 13 A
15 ft., SJT, 3 cond., 15 A
4.6 m, HAR, 3 cond., 10 A
Device ratings
110-120 VAC, 50/60 Hz, 11.7 A, or
220-240 VAC, 50/60 Hz, 6 A, or
100 VAC, 50/60 Hz, 12 A
Fuses
Type: Fast acting ceramic fuses, 250 VAC
AMP rating:
Operating speed: F (Fast Acting)
Breaking capacity:
3M™ Bair Hugger Model 675 Warming Unit – Service Manual
12 A (110 - 120 VAC)
8 A (220 - 240 VAC)
15 A (100 VAC)
15 A, 12 A: 750 A @ 250 Vac
15 A, 12 A: 10,000 A @ 125 Vac
8 A: 200 A @ 250 Vac
8 A: 10,000 A @ 125 Vac