Mechanical installation and wiring
Fig. 3:
Fig. 4:
Unscrew the two Allen screws and screw two spacer bolts into the threaded holes pro-
vided. See
Before attaching the mounting plate, attach the 4 circuit board holders with 4 screws
and 4 washers. See
Attach the mounting plate to the two spacer bolts with screws, washers, and lock wash-
ers. See
Clip the D31 board onto the circuit board holders. Make sure the board clicks in all the
way in the circuit board holders. Note the correct positioning of the board here:
(Fig. 6 / p.

Wiring

Removing the wire from the RNH.3 and RNH.4 relays to D1.X14
At the RNH (emergency stop relay), remove the RNH.3 and RNH.4 wires.
On the D1 board, detach the D1.X14 plug.