8
2-

ARCADIS Varic

2Mechanical installation and wiring

Opening the covers

Open the rear cover panel of the basic unit.
Disconnect the ground wire from the cover panel.

Installing the new footswitch holder

Once the rear cover panel has been removed from the basic unit, you can see three
holes in it that are fitted with dummy covers.
Remove the three dummy covers from the cover panel.
Attach the new multi-function footswitch holder to the cover panel using the screws,
lock washers, washers, and nuts included in the accessory kit.
NOTE

Installing the D31 board

Fig. 2:
ARCADIS Varic/Orbic
Mechanical installation and wiring
The holder for the normal footswitch, mounted on the front of the
basic unit, can remain on the basic unit.
If the customer would like this holder to be removed, unscrew it
and close the two threaded holes with the cover panel screws
included in the accessory kit.
D31 holder, installation location
SPR2-310.812.03.01.02
09.06
Page 8 of 20
CS PS SP
0
0
0
0
Siemens AG
Medical Solutions