Gxm640 Emi, Shock And Vibration Certifications; Table 4: Gxm 640 Emi, Shock And Vibration Certifications - Dalsa Calibir GXM640 User Manual

Table of Contents

Advertisement

The information contained in this document is not subject to Export Control Regulation in Canada.

GXM640 EMI, Shock and Vibration Certifications

Table 4: GXM 640 EMI, Shock and Vibration Certifications

Test Name / Standard
Measurement of conducted emissions
CISPR 11: 2015 A1: 2016 A2: 2019
Measurement of radiated emissions
CISPR 11: 2015 A1: 2016 A2: 2019
Measurement of conducted emissions - LAN port
CISPR 32: 2015
Measurement of conducted emissions
FCC Part 15: 2018, Subpart B
Measurement of radiated emissions
FCC Part 15: 2018, Subpart B
Radiated electromagnetic field immunity – radio frequencies
IEC 61000-4-3: 2006 A1 : 2007 A2 : 2010
Conducted immunity
IEC 61000-4-6: 2013
Electrostatic discharge immunity
IEC 61000-4-2: 2008
Electrostatic fast transient immunity
IEC 61000-4-4: 2012
Surge immunity
IEC 61000-4-5: 2014 A1: 2017
Magnetic field immunity
IEC 61000-4-8: 2009
Voltage dips, short interruptions and voltage variation immunity
IEC 61000-4-11: 2004 A1: 2017
RoHS
For an image of Calibir certificates see the section EC & FCC Declarations of Conformity.
Vibration & Shock Tests
Random vibrations
Shocks
Thermal Shock
Additional information concerning test conditions and methodologies is available on request.
Calibir GXM640 IR Camera User's Manual
Test Levels (while operating)
Level 1:
2 grms 60 min.
Level 2:
4 grms 45 min.
Level 3:
6 grms 30 min.
Level 4:
7.7 grms 60 min
Level 5:
15 grms 30 min
Level 6:
20 grms 30 min
Level 1:
20
g
Level 2:
30
g
Level 3:
40 g / 6 ms
-35
o
C - 65
o
C (+ 3
o
C)
Limit / Test Level
Group1, Class A
Group1, Class A
Class A
Class A
Class A
Scan
10 V/m 80-1000 MHz
3 V/m 1.4-2.7 GHz
Spot
3 V/m 1.8, 2.6, 3.5 and 5 GHz
10 V power
10 V I/O
±4 kV contact
±8 kV air
±2 kV power
±1 kV I/O
±1 kV L - L
±2 kV L - Ground
30 A/m / 50 Hz
0% - 1 cycle
40% - 10 cycles
70% - 25 cycles
0% - 250 cycles
Compliancy as per European directive 2011/65/EC
Test Parameters
Frequency range: 20 to 2000 Hz
Directions: X, Y, and Z axis
/
11
ms
Shape: half-sine
/
11
ms
Number: 6 shocks (+) and 6 shocks (-)
Directions: ±X, ±Y, and ±Z axis
50 cycles