The information contained in this document is not subject to Export Control Regulation in Canada.
Input Signal Electrical Specifications
External Input Block Diagram
Input 1 (pin 5)
Common Ground (pin 3)
External Input Details
Opto-coupled with internal current limit.
Single input trigger threshold level: <0.8V=Logical LOW, >2.4V=Logical HIGH.
Used as trigger acquisition event, counter or timestamp event, or integration
control.
User programmable debounce time from 0 to 16000 µs in 1µs steps.
Source signal requirements:
Single-ended driver meeting 3V, 12V, or 24V standards (see table below)
Differential signal driver
Calibir GXM640 IR Camera User's Manual
Current
Protection
Limiter
Technical Specifications 127