Dalsa Calibir GXM640 User Manual page 116

Table of Contents

Advertisement

The information contained in this document is not subject to Export Control Regulation in Canada.
An Example with two Calibir Cameras
The following basic steps configure two Calibir cameras connected to one computer via
an Ethernet switch, configured with two instances of CamExpert, to grab a frame every
second, controlled by a modulo event via PTP.
For each camera set features as follows:
I/O Controls — select Trigger Mode=ON, Trigger Source=Timestamp Modulo Event
Event Controls — select PTP Mode=Automatic
Note how one Calibir automatically becomes Master while the other becomes Slave
Event Controls — to have a modulo event every second, set Timestamp Modulo
Event=1000000000 (nanoseconds)
Click Grab on each instance of CamExpert. With the two cameras aimed at the same
moving object, you see that each camera grabs a frame at the same time.
IEEE 1588 Reference Resources
For additional information:
http://standards.ieee.org
PTP Standard Reference: IEEE Std 1588-2008 — IEEE Standard for a Precision Clock
Synchronization Protocol for Networked Measurement and Control Systems
116  Operational Reference
Calibir GXM640 IR Camera User's Manual