Using Dx-D 45C, Dx-D 45G Only Outside The Bucky - AGFA 5523/100 User Manual

Bucky
Table of Contents

Advertisement

40 | Bucky | Bucky
Orientation in the radiographic wall stand left loading
• To use the detector in portrait orientation, insert the detector in portrait
mode with the cable at upper left hand side.
• To use the detector in landscape orientation:
1. Insert the detector in portrait mode with the cable at upper left hand
2. Rotate the detector in the bucky.
Figure 17: Landscape orientation in the radiographic wall stand left loading
Orientation in the radiographic wall stand right loading
• To use the detector portrait orientation, insert the detector in portrait
mode with the cable at bottom right hand side.
• To use the detector in landscape orientation:
1. Insert the detector in portrait mode with the cable at bottom right
2. Rotate the detector in the bucky.
Figure 18: Landscape orientation in the radiographic wall stand right loading

Using DX-D 45C, DX-D 45G only outside the bucky

Use the DX-D 45C and DX-D 45G detector only for free exposures. Do not put
the DX-D 45C or DX-D 45G detector inside the bucky of the radiographic table
or the radiographic wall stand.
3253F EN 20210312 1125
side.
1
hand side.
1
2
2

Advertisement

Table of Contents
loading

Table of Contents