Typical Data; Specifications - JAI LQ-201CL User Manual

Rgb color & nir 4cmos line scan camera
Table of Contents

Advertisement

11.

Specifications

11.1 Typical data

Scanning system
Image Sensor
Pixel clock
Full
resolution
/Binning
Line Rate
Sub-sampling
/Windowing
Sensor sensitivity
Sensitivity
(Standard)
S/N
Synchronization
Video output
Gain range (Note1)
White balance
Black level
(User setup)
Line rate
(Variable) (Note2)
Electronic shutter
Binning
LQ-201CL
Line Scan
Effective pixels : 2048 pixels
Pixel Size : 14.0 μm × 14.0 μm
Effective image length : 28.672 mm
84.00 MHz
Total number of pixels per line: 2544 clocks
Line rate: 30.29 μs (No-shutter mode with internal trigger)
Line Frequency: 33.02 kHz
Total number of pixels per line: 1520 clocks
Line rate: 18.1 μs (No-shutter mode with internal trigger)
Line Frequency: 55.26 kHz
2
64nJ/cm
RGB : 2800 Lx ( 7800K White LED light)
Conditions: Line Rate=600μs, Gain=0dB, Shutter=OFF, Lens
iris=F2, 100% output)
2
NIR : 2.35μW/cm
at 800nm
55 dB (Green channel, Gain=0dB)
Internal
Digital 8-Bit x 4 or 10-Bit x 4 (Camera Link)
* Default setting: 8-bit
* Built-in pre emphasis function makes 10m transmission possible
Mode select:
Master mode : Adjust master level and match R,G,B and NIR
Master(Green): 0dB to +8dB
R/B/NIR : -4dB to +6dB
Individual
: Adjust each channel individually
R/G/B/NIR : -4dB to +14dB
* Default setting: Master mode
Adjustable range : 4000K to 9000K
Standard color temperature : 7800K
Mode select:
Master mode : Adjust master level and match R,G,B and NIR
Master(Green): 0 to 64 LSB(10-bit output)
R/B/NIR : -64 to 63 LSB (10-bit output)
Individual
: Adjust each channel individually
R/G/B/NIR : 0 to 64 LSB (10-bit output)
* Default setting: Master mode 8-bit (8 LSB)
Range : 30.29 μs (1L) to 15.0 ms (SensReadout is OFF)
Adjustment increment : 11.9 ns (1clk)
This mode is available for No-shutter internal trigger and
Shutter-Select internal trigger modes.
Available for Shutter-Select mode
Adjustable range : 9.52 μs to 14.99 ms
Adjustment increments : 11.9 ns (1 clk)
Available (line rate is not increased)
64

Advertisement

Table of Contents
loading

This manual is also suitable for:

Lq-201cl-f