PROBLEM
Improper
Delivery of
Probe Cover
or
Probe Cover
Positioned
Incorrectly
on Probe
Probe Cover
Breakage
SOLUTION
Probe covers positioned improperly within probe
cover dispenser. Remove dispenser from ther-
mometer handle. Pull on first probe cover to remove
any misaligned or bent covers. Re-position first
probe cover to extend 1/8" beyond top of dispenser,
then reinstall dispenser to thermometer handle. If
problem persists, replace probe cover dispenser.
Obstruction between probe and probe cover dis-
penser. Remove dispenser, inspect thermometer
handle and remove debris, if any. Ensure probe
covers are properly positioned in probe cover dis-
penser, then reinsert dispenser into handle.
Probe tip is damaged. Replace with new probe tip.
See "Maintenance" section of this manual for proce-
dure.
Damaged probe covers. Replace probe cover dis-
penser.
Probe tip is damaged. Replace with new probe tip.
See "Maintenance" section of this manual for proce-
dure.
Probe covers positioned improperly within probe
cover dispenser. Remove dispenser from ther-
mometer handle. Pull on first probe cover to remove
any misaligned or bent covers. Re-position first
probe cover to extend 1/8" beyond top of dispenser,
then reinstall dispenser to thermometer handle. If
problem persists, replace probe cover dispenser.
9020 Users' Manual
25