Chapter Xiii System Prompts And Trouble-Shooting; Definitions Of System Prompts; Troubleshooting - UNI-T UTD2052CL+ User Manual

Digital oscilloscope
Hide thumbs Also See for UTD2052CL+:
Table of Contents

Advertisement

Chapter XIII System Prompts and Trouble-

13.1 Definitions of System Prompts

Adjustment at Ultimate Limit: This informs you that the multifunction control knob has
reached its adjustment limit in the current status. No further adjustment is possible. When
the vertical deflection factor switch, time base switch, X shift, vertical shift and trigger level
adjustments have reached their ultimate limits, this prompt will appear.
USB Device Unplugged: After a USB key is unplugged from the oscilloscope, this prompt
appears.
USB Screen Shot: When pressing Prtsc, this prompt appears.
USB File Save: When the oscilloscope is saving a waveform, this prompt is shown on the
screen.
No signal in channel: When inputing slow signal, small signal or no signals, after
conducting auto setting, this prompt appears.

13.2 Troubleshooting

(1) No waveform appears
If no waveform appears on the screen after acquiring signals, please follow the steps below
to find the cause:
①. Check if the probe is connected to the signal test point correctly.
②. Check if the signal connection line is connected to the input point of analog channel.
③. Check if the input point of analog channel signal is consistent with the open channel.
④. Connect the probe tip to the compensating signal connector of DSO to check if the
probe is in good condition.
⑤. Check if the thing to be measured generates signals (connect the channel with signal
to the signal without signal to find the cause).
⑥. Press AUTO to recollect signals.
(2) Voltage test problem
The measured voltage amplitude is 10 times higher or lower than the actual value: check
if the attenuation coefficient of channel probe is consistent with the attenuation rate of
applied probe.
(3) No trigger
There is waveform display but it is not stable:
① Check the trigger Source setup in the trigger menu. See whether it is the same as the
actual signal input channel.
② Check the trigger type: Use Edge trigger for ordinary signals. Stable waveform display
is achieved only when the correct trigger mode is selected.
③ Try changing the trigger coupling to High Frequency Holdoff or Low Frequency Holdoff
shooting
57
UTD2000 Series User Manual