6. System Setup

This chapater includes the contents as follows.
Page Explaination of <SYSTEM 1>
Page Explaination of < SYSTEM 2>
Page Explaination of <System Information>
6.1 <System> Setup
<System Setup> page is to set some settings that are not related to the specific test item parameters, but are
related to the test scheme of the instrument.
When high voltage output is stop, press [System Setup] key to enter <SYSTEM 1> page.
Table 6-1 Explaination of <SYSTEM 1> Page
Item
Input Range
Start
Local,PLC
mode
Low,middle,high
Volume
Key sound
ON, OFF
Long and short sound,
Pass beep
double short sound, OFF
Long sound, double short
Fail beep
sound, OFF
(0.1~99.9)s,OFF
Start delay
Instruments.uni-trend.com
Figure 6-1 <System Setup 1> Page
Default Value
Description
Local: Press start key on the front panel to boot-up.
PLC: Input START control signal by HANDLER to
Local
boot-up
Communication: It'a always valid, the specific
refer to the communication protocol.
Middle
The volume setting of beeper.
ON
The switch of key sound.
Long
and
The sound mode of pass beeper.
short sound
Long sound
The sound mode of failed beeper.
OFF
The delay time from start to the first test item.
UT5300X+ and UT5320R-SxA Series
25 / 43