Linearisation with Direct Calibration - cont'd.
Set1
5
Digit
Max/Min
Set1
Digit
Max/Min
6
Set1
Digit
Max/Min
7
Display
rd 08
rd 07
rd 06
rd 05
rd 04
rd 03
rd 02
rd 01
rd 00
Set00
Set2
Output
Alarms
OK
Reset
Set2
Output
Alarms
OK
Reset
Set2
Output
Alarms
OK
Reset
Set01
The display will show
Apply your 1st known load to the system
and press OK.
The display will now show
You can now tell the meter what it should
display for this load.
Select each digit in turn with the DIGIT
button, and increase or decrease each
digit's value using the UP or DOWN
button, until the display is set as required.
Press OK when done.
The display will show
Press OK if you want to Add another
calibration point, you will repeat the previ-
ous steps, each time incrementing "Set
XX" and "rd XX" up to Set 09 and rd 09
which are the last available points.
If you want to finish adding points, press
Set2 until the display shows
then confirm by pressing OK
If you want to abort the setup, press Set2
until the display shows
press OK
Set02
Set03
Set04
SEt 01
rd 01
Add.Pt
, and
SAVE
quit
and then
Done!
Set05 Set06
Set07
Input
Set08
21