Operating Instructions
The patient is the intended user and can safely use all functions.
Preparing the Device for Use
• Remove garments and cardboard underneath from box
• Lift re-shipper box out of outer box and open
• Remove pumps and charger – save packaging for transport and storage
• Connect charger to pumps and plug into outlet
• Indicator light will be a steady yellow until pump is fully charged – a depleted battery
can take up to 4 hours to charge
• Assure pumps are fully charged before first use
• Remove garments from plastic bag, unroll, and spread flat
Connecting the Garments
• Firmly insert pump valve into garment input tube
• Attach pump to garment using Velcro on back
Putting the Garments On
• Snugly wrap sleeve marked "L" around left calf and sleeve marked "R" around right
calf – secure using Velcro fasteners
• Ensure air chamber is positioned at back of calf
Operating the Device
• Press "Power On/Off" button to turn on
• Indicator light will flash green – if it simultaneously flashes yellow, the battery is low
and must be charged
• Pump will inflate garment and hold for 15 seconds every minute until turned off with
power button
• If sleeve does not properly inflate in 3 minutes, pump will stop operating and alarm will
sound until power button is pressed or battery dies
• Should battery become low, an alarm will sound for 1 second each minute and
indicator will turn yellow
• Upon completion of treatment, press "Power On/Off" button to turn pump off
• Charge pumps for next use
• Upon the completion of treatment or to stop treatment, press "Power On/Off" button to
turn off
Reading the Usage Meter
Begin with device turned off. Press and hold the "Power On/Off" button for 10 seconds
until you hear a short beep. Release button and you will hear a long beep. The pump
7
Need help?
Do you have a question about the Vascu-Ease IC-1200-WH and is the answer not in the manual?
Questions and answers