Notes on trouble-shooting

Checking the kV and tube current (IR)

Connect oscilloscope on D910 to MP "kV", "IR" and "GND".
POLYMOBIL ON
Trigger exposure with the default values 60 kV, 10 mAs.
kV
IR
kV
IR
Prolonged transient response
NOTICE
Siemens AG
Medical Solutions
The transient response of the tube current can be set with P1 on
D910.
RXB8-115.061.02
Rev. 01