Notes on trouble-shooting
Checking the kV and tube current (IR)
•
Connect oscilloscope on D910 to MP "kV", "IR" and "GND".
•
POLYMOBIL ON
•
Trigger exposure with the default values 60 kV, 10 mAs.
kV
IR
kV
IR
Prolonged transient response
NOTICE