Notes on trouble-shooting
•
Measure the supply voltages on D910.X3.
+ 5 V at X3.4 and X3.1 (0 V)
+15 V at X3.5 and X3.6 (0 V)
- 15 V at X3.7 and X3.6 (0 V)
Siemens AG
Medical Solutions
Tolerance
Tolerance
Tolerance
RXB8-115.061.02
Rev. 01
10.95
D910 (X038I-91E)
±
0.5 V
±
1 V
±
1 V
Page 7 of 24
CS PS 24
4 - 7
POLYMOBIL III