For example, in
Figure
syntax:
Range 30V high (+30 V)
980000
5. Points where the calibration procedure asks for no value. Find the calibration value by any
suitable method or use the value from a calibration certificate. Use the cursor keys or the
rotary knob to adjust the value to write down.
These functions use this system:
HV Probes
High Resistance Multiplier
17, the display identifies the active resistance segment in this
DC voltage range 30 V of internal voltage calibrator,
nominal value +30 V dc
Adjustable calibration value. Set the value to get the
exact output voltage +30.000 V on a standard
voltmeter
Figure 17. DCV Screen
Multifunction Electrical Tester Calibrator
Calibration
19