Making Measurements
When Using CONFigure
Note
Chapter 4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
To make a measurement the multimeter must be in the wait-for-trigger state
when a trigger signal occurs. The
the multimeter in the "wait state" after setting the configuration. When
is used, the multimeter must be placed in the wait state with the
CONFigure
command:
(readings are sent to the output buffer)
READ?
or
INIT[:IMMediate]
These commands follow
CONFigure
CONF:FRES 1861,MAX,(@100:103)
RES:OCOM ON
TRIG:COUN 3
READ?
CONF:FRES 1861,MAX,(@100:103)
RES:OCOM ON
TRIG:COUN 3
INIT
READ?
and
INIT
will make measurements upon execution if
(trigger signal always true) remains set. If the trigger source is changed
following the
command, execution of the
CONFigure
commands will place the multimeter in the wait-for-trigger state; however, a
measurement will not be made until a trigger from the specified source occurs.
Understanding the HP E1326B/E1411B Multimeter
command automatically places
MEASure
(readings are stored in memory)
as shown below:
READ?
TRIG:SOUR IMM
or
INIT