Subsystem Syntax

:LFRequency

Parameters
Comments
Example
:LFRequency?
123
Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The

CALibration

command subsystem selects the multimeter's line reference
frequency (
CALibration:LFRequency
(
CALibration:ZERO:AUTO
CALibration
:LFRequency <frequency>
:LFRequency? [MIN | MAX]
:ZERO:AUTO <mode>
:ZERO:AUTO?
CALibration:LFRequency <
by the multimeter's analog-to-digital converter.
Parameter Name
<
frequency
>
• MIN
selects the minimum line reference frequency (50 Hz).
maximum line reference frequency (60 Hz).
The line reference frequency is set to 60 Hz at the factory. The setting is stored in
non-volatile mainframe memory and is changed only when
For a line frequency of 400 Hz, the 50 Hz reference frequency is used; however,
since 50 Hz is a subharmonic of 400 Hz, it provides normal mode rejection of
power line related noise.
*RST Condition: The selected line reference frequency remains unchanged
since it is stored in non-volatile mainframe memory.
Selecting the Line Frequency Reference
CAL:LFR 50
CALibration:LFRequency? [MIN | MAX]
the output buffer:
The present line reference frequency ("50" or "60") if
not specified.
The minimum line reference frequency ("50") if
The maximum line reference frequency ("60") if
) and enables/disables the autozero mode
).
frequency
>
selects the line reference frequency used
Parameter Type
Range of Values
numeric
50 | 60 | MIN | MAX
!Reference frequency is 50 Hz.
returns one of the following numbers to
MIN
MAX
CALibration
Default Units
hertz
MAX
selects the
is executed.
CAL:LFR
MINimum
or
MAXimum
is
is specified.
is specified.
Chapter 5