Resistance:aperture; Resistance:nplc - HP E1326B User Manual

5 1/2 digit multimeter
Table of Contents

Advertisement

RESistance:APERture?

Example

RESistance:NPLC

Parameters
Comments
Example
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
[SENSe:]RESistance:APERture? [MIN | MAX]
numbers to the output buffer:
The present aperture time in seconds if
The minimum aperture time available (10 µs) if
The maximum aperture time available (320 ms) if
Querying the Aperture Time
RES:APER 2.67E-01
RES:APER?
enter statement
[SENSe:]RESistance:NPLC <
cycles (PLCs). Values are rounded up to the nearest number of PLCs shown in the
following table.
Parameter Name
<
number
>
• MINimum
selects 0.0005 PLCs.
integration time in power line cycles (PLCs) also sets the aperture time and the
resolution. For example, 16 PLCs (60 Hz line frequency) sets an aperture time of
267 ms. The corresponding resolution depends on the function and range you
select.
The
command overrides the results of previously executed
RES:NPLC
RESistance:APERture
and
executed has priority).
The greater the number of PLCs, the greater the normal mode rejection (and the
lower the reading rate).
Only the 1 PLC and 16 PLC settings provide normal mode rejection of 50 Hz or
60 Hz power line related noise.
*RST Condition: 1 PLC
Setting the Integration Time in PLCs
RES:NPLC 16
returns one of the following
or
is not specified.
MIN
MAX
MIN
MAX
!Aperture time is 267 ms.
!Query multimeter to return aperture time.
!Enter value into computer.
number
>
sets the integration time in power line
Parameter Type
Range of Values
numeric
0.0005 | 0.005 | 0.125 | 1 | 16 |
MIN | MAX
MAXimum
selects 16 PLCs. Setting the
RESistance:RESolution
!Integration time is 16 PLCs.
Command Reference
is specified.
is specified.
Default Units
PLCs
commands (the last command
164