Format; [:Data] - HP E1326B User Manual

5 1/2 digit multimeter
Table of Contents

Advertisement

Subsystem Syntax

[:DATA]

Parameters
Comments
Example
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The

FORMat

subsystem sets the format for data transferred from the multimeter to
the computer using the
MEASure?
FORMat
[:DATA] <type>[,<length>]
type
FORMat[:DATA] <
>[,<
Parameter Name
<
type
>
length
<
>
ASCII readings are transferred in the form ± 1.234567E ± 123. Each reading is
followed by a comma(,). A line feed (LF) and End-Or-
Identify (EOI) follow the last reading. Each reading is 15 bytes.
REAL reading lengths are 32- and 64-bits. The readings are transferred in the
IEEE 488.2-1987 Definite Length Arbitrary Block format. The readings are
preceded by an Arbitrary Block header which consists of:
# <non-zero digit> <block length>
<non-zero digit> indicates the number of digits representing
<block length>, and <block length> indicates the number of 8-bit data bytes
which follow the header. Examples of the header are shown below:
REAL,32
#240
<40 bytes>
REAL,64
#18
<8 bytes>
*RST Condition: ASCii
Setting the Data Format
FORMAT REAL,64
CONF:VOLT:DC 58.1
,
READ?
, and
FETCh?
length
selects the data format and length.
>]
Parameter Type
Range of Values
discrete
numeric
10 readings
1 reading
!Format is set to REAL 64.
Command Reference
FORMat
commands.
Default Units
ASCii | REAL
none
32 | 64
none
140