Maximizing Measurement Accuracy

Comments
Chapter 3
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
This program makes DC voltage measurements on three channels using the
multimeter configuration which makes the most accurate measurements.
Note that measurement accuracy also depends on wiring practices and the
surrounding environment.
10
! Dimension a computer array to store the readings.
20
DIM Rdgs(1:3)
30
! Clear and reset the multimeter.
40
CLEAR 70903
50
OUTPUT 70903;"*RST"
60
! Configure the multimeter for DC voltage measurements.
70
! Set autorange and select the minimum (best) resolution.
80
OUTPUT 70903;"CONF:VOLT:DC AUTO,MIN,(@101:103)"
90
OUTPUT 70903;"READ?"
100
! Enter and display the readings on the computer.
110
ENTER 70903;Rdgs(*)
120
PRINT Rdgs(*)
130
END
sets the minimum (best) resolution for the range set by
MIN
autorange.
also indirectly selects the aperture time for the most
MIN
accurate measurements.
The terms
MIN
and
command's syntax.
parameter.
selects the maximum numeric value for the
MAX
parameter. This eliminates the need to look up specific numbers in
the manual.
turns autozero on.
CONFigure
When making resistance measurements (including thermistors and
RTDs), accuracy can often be increased by turning on offset
compensation (for example,
Using the HP E1326B/E1411B Multimeter
MAX
often appear as parameter choices in a
selects the minimum numeric value for that
MIN
RESistance:OCOMpensated
).