The following program resets the multimeter.
10
! Map the A16 address space in the HP V/382 and store the multimeter base
20
! address in a variable.
30
CONTROL 16,25;2
40
COM Base_addr
50
Base_addr=DVAL("C600",16)
60
! Call the subprogram which resets the multimeter.
70
CALL Mm_rst
80
END
90
! This subprogram checks each bit in the multimeter Status register.
100
! The subprogram is called by subprogram Mm_rst to monitor status bit 2.
110
SUB Read_status(Status,Rdy,Done,Indardy,Qryrdy,Noerr,Pass_fail)
120
COM Base_addr
130
Status=READIO(-16,Base_addr+4)
140
Rdy=BIT(Status,0)
150
Done=BIT(Status,7) AND Rdy
160
Indardy=BIT(Status,4) AND Done
170
Qryrdy=BIT(Status,1) AND Done
180
Noerr=NOT (NOT (BIT(Status,6)) AND Done)
190
Pass_fail=BIT(Status,2)
200
SUBEND
210
! This subprogram resets the multimeter by disabling the SYSFAIL bit,
220
! then writing a '1' to Control Register bit 0, and then writing a '0' to
230
! Control Register bit 0. Once the reset completes, SYSFAIL is re-enabled.
240
SUB Mm_rst
250
COM Base_addr
260
WRITEIO -16,Base_addr+4;
270
WRITEIO -16,Base_addr+4;3
280
WRITEIO -16,Base_addr+4;2
290
REPEAT
300
CALL Read_status(Status,Rdy,Done,Indardy,
Qryrdy,Noerr,Pass_fail)
310
UNTIL Pass_fail
320
WRITEIO -16,Base_addr+4;0
330
SUBEND