Using a Single
Trigger
Example:
Measurements with a
Single Trigger
Aborting a
Measurement
112
Understanding the HP E1326B/E1411B Multimeter
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
The multimeter can be internally triggered with a single trigger signal. This
signal is issued with the
TRIGger[:IMMediate]
Before a single trigger can be sent, however, the trigger source must be set
to
and the multimeter must be in the wait-for-trigger state. These
HOLD
conditions are shown in the following program which shows how a single
trigger can be used.
When
TRIG
is executed, the multimeter makes 10 DC voltage
measurements and stores them in mainframe memory. The
command retrieves the readings.
10
! Configure the stand-alone multimeter for DC voltage measurements.
20
! Suspend triggering and set 10 measurements to be taken when the single
30
! trigger is received.
40
CONF:VOLT:DC
50
TRIG:SOUR HOLD
60
SAMP:COUN 10
70
! Place the multimeter in the wait-for-trigger state and issue a single trigger.
80
! Fetch the readings from memory.
90
INIT
100
TRIG
110
FETCH?
When the multimeter is in the wait-for-trigger state it can be returned to the
idle state before the trigger signal is received. This is done with the
command.
Trigger sources which allow the
to the idle state are
TRIGger:SOURce BUS
The next two examples show how
command:
ABORt
command to return the multimeter
and
TRIGger:SOURce HOLD
ABORt
works.
FETCH?
ABORt
.