Examples

:COUNt?

Example
Chapter 5
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
• CONFigure
and
MEASure
*RST Condition:
TRIG:COUN 1
Setting the Trigger Count (Scanning Multimeter)
CONF:VOLT:DC (@100:104)
TRIG:SOUR EXT
TRIG:COUN 10
READ?
enter statement
Setting the Trigger Count (Stand-Alone Multimeter)
CONF:VOLT:DC
TRIG:SOUR EXT
TRIG:COUN 10
READ?
enter statement
TRIGger:COUNt? [MIN | MAX]
buffer:
The present trigger count (1 through 16,777,215) if
The minimum trigger count available (1) if
The maximum trigger count available (16,777,215) if
Querying the Trigger Count
TRIG:COUN 10
TRIG:COUN?
enter statement
set the trigger count to 1.
!Function: DC voltage;
specify channel list.
!Trigger source is external BNC on
multimeter front panel.
!Multimeter will accept 10 external triggers
(5 channels will be scanned with each
trigger).
!Place multimeter in wait-for-trigger state;
make measurement when external trigger is
received; send readings to output buffer.
!Enter readings into computer.
!Function: DC voltage; stand-alone
multimeter.
!Trigger source is external BNC on
multimeter front panel.
!Multimeter will accept 10 external triggers
(one measurement is taken with each
trigger).
!Place multimeter in wait-for-trigger state;
make measurement when external trigger is
received; send readings to output buffer.
!Enter readings into computer.
returns one of the following numbers to the output
MIN
MIN
is specified.
MAX
!Multimeter will accept 10 external triggers.
!Query multimeter to return trigger count.
!Enter value into computer.
Command Reference
or
are not specified.
MAX
is specified.
180