Download Print this page

Technical Data - Integra PIPETBOY acu 2 Operating Instructions Manual

Hide thumbs Also See for PIPETBOY acu 2:

Advertisement

PIPETBOY acu 2 - Operating instructions V17
5.3
Equipment disposal
PIPETBOY acu 2 device must not be disposed of with unsorted municipal
waste. Do not dispose of the device in a fire.
PIPETBOY acu 2 contains a Li-ion battery. Do not modify the battery in any
way. Dispose of the PIPETBOY acu 2 device and the battery separately in
accordance with the laws and regulations in your area governing disposal of
devices containing Li-ion batteries.
In certain regions and countries, e.g. in all EU member states, the distributor is obliged to
take back this product free of charge at the end of life. Please contact your local distributor
for more details.
6

Technical Data

6.1
Specifications
Pipetting speed
Battery
Electricity supply
Materials
Dimensions (H x W x D)
Weight
Ambient conditions
14
max. 13.5 ml/s (with a 50 ml serological pipette)
Type: rechargeable, Li-ion, min. 500 mAh
Typical charging time: 3.5 hours
Charging cycles: 500–1000 (when charging as indicated)
Running time: at least 5500 cycles of aspiration and
dispensing of 25 ml.
Mains adapter input: 100–240 VAC, 50/60 Hz
Device Input: 16–19 VDC, 3.1 W
Housing: PA
Nose piece housing: POM
Pipette mount: Silicone
Filter rubber: Silicone
125 x 130 x 35 mm
195 g
Operation: 5–40°C, max. 80% RH
Storage: -10–50°C, max. 95% RH
Technical Data
INTEGRA Biosciences AG