To Pump Condensable Vapours; To Decontaminate The Oil; Unattended Operation; Shut-Down - Edwards A373-30-940 Instruction Manual

Rotary vacuum pumps
Table of Contents

Advertisement

A373-10-880 Issue R
4.5

To pump condensable vapours

Use gas-ballast (open the gas-ballast control) when there is a high proportion of condensable vapours in the process
gases:
1. Close the vacuum system isolation-valve.
2. Turn the gas-ballast control anti-clockwise to fully open and operate the pump for 30 minutes to warm the oil;
this will help to prevent vapour condensation in the pump.
3. Open the vacuum system isolation-valve and continue to operate the pump with the gas-ballast control open.
After you have pumped condensable vapours, you can (if necessary) decontaminate the oil; use the procedure in
Section
4.6.
4.6

To decontaminate the oil

The oil in the pump should be clear, if the oil is cloudy or discoloured, it is contaminated with process vapours.
1. Look at the condition of the oil in the oil sight-glass
continue with the procedure at
2. Close the vacuum system isolation-valve.
3. Turn the gas-ballast control fully anti-clockwise.
4. Operate the pump until the oil is clear.
4.7

Unattended operation

The pump is designed for unattended operation under the normal operating conditions specified in
However, we recommend that you check the pump at a regular interval of not more than 14 days; check the pump
more frequently if you pump high volumes of gas or if you operate the pump with the gas-ballast control open.
Single-phase motors are cooled by internal fans. These motors have a thermal overload device. When the motor is
too hot, the thermal overload device switches off the pump. The thermal overload device has an automatic reset;
when the motor cools down, the device resets and the motor will restart.
When you check the pump, make sure that the pump is not going through a repetitive cycle of thermal overload
failures and automatic resets. If necessary reduce the thermal load from the pumped gases, to prevent overheating
of the pump.
4.8

Shut-down

Note:
If the gas-ballast control is open and the motor is switched off for any reason, the pump drive shaft may
rotate in the reverse direction, causing a system pressure rise. To prevent this, use a gas-ballast control
valve (refer to
Section
We recommend, as described in the procedure below, that you decontaminate the oil before you shut down the
pump; this will prevent damage to the pump by the contaminates in the oil.
1. Refer to
Section 4.6
and decontaminate the oil, as required.
2. Close the vacuum system isolation-valve (if not already closed).
3. Close gas-ballast (that is, turn the gas-ballast control clockwise).
4. Switch off the electrical supply to the pump.
Page 22
(Figure
Step 2
below.
7.5.15).