8
AE Area Display Setting
}The average or spot metering area can be displayed on or hidden from
the monitor screen.
See page 27 for the metering area setting.
9
Measurement Functions
Distances and areas in the image can be measured. The measure-
ments are possible by using the hand switch buttons or the mouse
connected to the control box D21-CB.
}The scale should be set before measurements. For the setting method, see Scale Display Setting (p. 38).
}The measurement data can be displayed either on the live image alone or on both the live and recorded images. For the
selection method, see Scale Display Range Setting (p. 40).
}We recommend you to use the same live resolution as the image quality mode during recording.
If the live resolution is different from the image quality mode, the measurement result may be shifted.
}If text input is required, connect the keyboard to the control box D21-CB.
}The measurement functions are simplified functions based on the CCD pixels and are not suitable for precise measure-
ments. If precise measurements are required, use appropriate measuring instruments.
}The measurement result is displayed to the fourth place of significant figures with the distance, to the fifth places of
significant figures with the angle, and to the seventh places of significant figures with the area. The unit used in the
measurement result display cannot be changed.
}The display result is displayed near the drawn line or circle. It is not possible to change the display position, the font, the
line thickness and the number of displayed digits.
1. Select [START MEASUREMENT]. The MENU display is switched to the measurement menu shown below.
2. Select the desired measurement from the measurement menu. The pointer on the screen turns into "
3. Move the pointers to the measurement target positions using the cursor buttons of the hand switch D21-HS, and press
the Shift button to view the measurement result on the screen. The diagram for the measurement method and the results
are displayed automatically. The digits of the measurement results are also set automatically. Now the next measurement
can be started (up to 20 measurement results, including those of previous measurements, can be displayed on the
screen).
4. To end measurements and return to the measurement menu display press the SET/OK button or MENU button (this is
also possible by right-clicking of the mouse if this is used). To cancel a measurement in the middle, press the MENU
button (or left-click on [END (MENU)] at the top right of the screen if you are using the mouse).
5. To return to the normal MENU display, display the measurement menu by ending measurement, and press the MENU
button of the hand switch D21-HS (or click on
Measurement menu
Downloaded from
ManualsNet.com
search engine
on the upper right of the MENU display if you are using the mouse).
DP21-SAL
Default: OFF
Display AE area
Hide
".
42