HCEN-24 Economical Capillary Centrifuge
PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION
Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529
mrc@mrclab.com
MRC.2.22

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the HCEN-24 and is the answer not in the manual?

Questions and answers

Summary of Contents for MRC HCEN-24

  • Page 1 User Manual HCEN-24 Economical Capillary Centrifuge PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION Hagavish st. Israel 58817 Tel: 972 3 5595252, Fax: 972 3 5594529 mrc@mrclab.com MRC.2.22...
  • Page 2: Table Of Contents

    Contents Safety Precautions ............................0 1. Specifications ............................. 0 2. Scope of application ........................... 0 3. Environmental conditions .......................... 0 3.1 Basic operating conditions ....................... 0 3.2 Transport and storage conditions ...................... 0 4.1 Mounting position ..........................1 4.2 Connection between power cable and ground wire ................1 5.
  • Page 3: Safety Precautions

    Safety Precautions Carefully read the following safety precautions for a thorough understanding. • The symbol is an internationally accepted safety mark. • Ensure to operate this centrifuge safely as per the instructions and procedures described in this manual. • Carefully read all safety messages in this manual for safe operation of this centrifuge. •...
  • Page 4 or pathogenic materials, clean it according to the decontamination procedures specified. ⚫ If you require service at site, please sterilize and decontaminate the centrifuge in advance, and then notify the service center about the details of the materials and procedure. ⚫...
  • Page 5 (3) The centrifuge tubes should be within the centrifuge actual capacity. (4) If the rotor has a cover, be sure to tighten the cover before operation. (5) If the rotor is provided with a lid, ensure it is secure it tightly before operation. ⚫...
  • Page 6: Specifications

    1. Specifications Maximum rotation speed 12,000rpm (500-12,000rpm) step: 100rpm Maximum RCF 13680× g, step: 100× g Capacity AC24 (capillary× 24) 10 s-99m-HOLD(continuous operation) Timing Drive motor DC brushless motor Safety performance Dual door locks, over speed, over temperature and status diagnosis system. Single phase,220-240V, 50Hz/60Hz,380W Power Dimensions(mm)
  • Page 7: Mounting Position

    4. Installation and testing Users must strictly comply with the installation instructions contained in this chapter. Be advised! Warning ⚫ Improper power connection might damage the centrifuge. ⚫ Ensure to check the power supply before connection for compliance with the requirements. 4.1 Mounting position (1) Place this centrifuge on a firm、flat and level surface, ensure the four feet of this centrifuge stand on the counter firmly.
  • Page 8: Operating Panel

    Upper cover Sight glass Vent Power socket Power switch Fig. 5.2 Rear view of centrifuge 6. Operating panel Speed display Door lock display Time display x 100rpm Fig. 6.1 Schematics of operating panel...
  • Page 9: Rotor Preparation

    Legend Name Function Select Button Press this button to choose the parameters setting. When the centrifuge lid is locked tightly, Press and hold this Pulse Button button and the centrifuge run to the set rotation speed Set the parameters to desired setting and press this button to Run start / stop key start run and operate the centrifuge.
  • Page 10: Ensure The Balanced Centrifuge Tube

    ⚫ Do not exceed the actual tube capacity as mentioned in this user manual. 7.3 Ensure the balanced centrifuge tube ⚫ Although the centrifuge may be used with visually confirmed balance, it is recommended that the samples are weighed using a balance to keep this centrifuge in a well-balanced condition to extend its life expectancy.
  • Page 11 Sec/ X 100rpm . Fig. 8-1 Startup interface of centrifuge .The centrifuge shows the centrifuge model HCEN-24, as shown in Fig. 8-2 Fig. 8-2 Model HCEN-24 interface The centrifuge displays the last operating parameters, as shown in Fig. 8-3. x 100rpm Fig.
  • Page 12 Caution ⚫ Place the rotor on the main shaft and ensure full contact between the rotor and the main shaft. Screw up the nut on the rotor tightly using a wrench to connect the rotor with the main shaft firmly, otherwise the rotor might fall off, causing damage to the centrifuge.
  • Page 13: Rcf Operation

    8.1.3 Start operation (1) Press the run start/stop button to start operation. • The upper lid is locked and the rotor starts rotating. • The timer starts only after the rotation speed reaches the set value, with the remaining run time display.
  • Page 14: Transient Operation

    • Press the select button to choose the rotation speed unit as xg. Relative centrifugal force enters into the adjustment mode if the centrifugal acceleration value flashes. • The centrifuge enters into operation mode if there is no activity for about 7Seconds •...
  • Page 15: Wearing Parts

    (2) Centrifugal chamber Caution • Never directly pour water or other solvents into the centrifugal chamber as these solvents might enter into the drive unit and cause corrosion or damage to the bearings. • If the centrifuge chamber becomes dirty, clean it using a cloth/ sponge soaked with neutral cleaning agent.
  • Page 16: Common Failures And Solutions

    reliable grounding. In case of disconnection, identify the cause and eliminate the failure before the centrifuge operation. 10. Common failures and solutions 10.1 List of common failures This centrifuge is capable of self diagnosis. When the centrifuge fails, the time display window will indicate the failure code, leading to the immediate identification of the cause of failure.
  • Page 17: How To Open The Upper Cover Lid

    10.2 How to open the upper cover lid Reminder: • Open the upper cover only when the rotor is not running, if the centrifuge is powered ON, (1) When the centrifuge is turned ON, the rotor stops running and the upper cover lock may be opened by pressing the door lock button on the right side of the machine.
  • Page 18: Replace The Fuse

    10.3 Replace the fuse (1) The centrifuge’s fuse is 250V, 5A time delay size: Ф5×20. (2) The centrifuge’s fuse is on the power socket. The fuse may be replaced by taking the fuse box out of the power socket. 11. Introduction to rotor and centrifuge tube Caution •...
  • Page 19 Maximum Maximum Maximum relative Rotor Rotor Centrifuge tube speed centrifugal force centrifugal radius number name (rpm) Rcf (× g) (cm) 24-ø 1.75× 75 AC24 12000 13680 capillary Table 11.1 Rotor 11.1.3 Precautions (1) The centrifuge rotor can separate samples with a density lower than 2.0g/ml. If the samples density is over 2.0g/ml, please calculate allowable speed depending on the following formula.
  • Page 20: Rcf Calculation

    11.2.2 PC centrifuge tube cleaning PC material has relatively low chemical stability to alkaline solvent, therefore use of cleaning agent with pH value of over 9 should be avoided. Some neutral cleaning agents still have pH value of over 9 after being diluted as recommended by the vendor, therefore use of cleaning agent with pH value of 7-9 is recommended.
  • Page 21: Ordering Information

    Capillary rotor package, including rotor, rotor cover and reading dial, 19400047 AC24 with maximum rotation speed of 12,000 rpm, maximum capacity of 24 capillaries, suitable for HCEN-24. 14. Warranty 14.1 Unit warranty The entire use will have two-year warranty period commencing from delivery date under the conditions of normal maintenance.