Analyzer Disassembly - Keysight N9032B PXA Manual

Table of Contents

Advertisement

Option EXM, External Mixing Upgrade Kit

Analyzer Disassembly

1. Disconnect the instrument from ac power.
2. Refer to
Figure
1. Using the T-20 driver, remove the 4 screws (1) (two on each side) that
attach the handle strap (2) on each side of the instrument.
3. Remove the four key locks from the instrument four bottom feet.
4. Remove the four instrument bottom feet (6).
5. Using the T-20 driver, remove the four screws (including washers) (3) that hold the rear feet
(4) in place.
6. Pull the instrument cover (5) off towards the rear of the instrument.
Figure 1
Instrument Outer Case Removal
6
If the instrument is placed on its face during any of the following procedures, be sure to use a soft
surface or soft cloth to avoid damage to the front panel, keys, or input connector.
Make sure any adapters on the front panel are removed.
Installation Note N9032-90004