Option EXM, External Mixing Upgrade Kit
Utilities, Adjustments, and Performance Verification Tests
Calibration Software and specified test equipment is required to perform the adjustments, and can
be used to automate the performance verification testing. Information on how to obtain this
software can be found at:
http://www.keysight.com/find/calibrationsoftware
Utilities Required
None
Adjustments Required
The following adjustments are the minimum set required to ensure that the newly installed
hardware is functioning properly
— IF Input Gain Adjustment
— LO Output Power Adjustment
Performance Testing Required
The following performance verification tests are the minimum set required to ensure that this newly
installed option is functioning properly.
— IF Input Gain Accuracy
— LO Output Power Accuracy
A full calibration is required to assure the instrument meets all specifications
The end user must ultimately determine whether they want a full calibration to be performed after
the installation of this upgrade or not. If a full calibration is required, arrangements regarding the
level of calibration must be made between the end user and the calibration provider.
For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your
local Keysight office access the following URL, or if in the United States, call the following
telephone number:
http://www.keysight.com/find/assist
1-800-829-4444 (8 am - 8 pm ET, Monday - Friday)
Installation Note N9032-90004
23