Technical data
15.1
Input
15.1.1
Measured variable
Level (point level), MAX or MIN safety
15.1.2
Measuring range
Depends on the installation location and the pipe extension ordered
Standard pipe extension up to 3 m (9.8 ft) and up to 6 m (20 ft) on request.
15.2
Output
15.2.1
Output signal
SIO (FEL60H electronic insert)
8/16 mA (SIO) with superimposed digital communication protocol HART, 2-wire
Continuous operation (FEL60H electronic insert)
4 to 20 mA proportional to the oscillation frequency with superimposed digital
communication protocol HART, 2-wire
For continuous current output, one of the following modes of operation can be selected:
• 4.0 to 20.5 mA
• NAMUR NE 43: 3.8 to 20.5 mA (factory setting)
• US mode: 3.9 to 20.8 mA
15.2.2
Signal on alarm
Signal on alarm in accordance with NAMUR recommendation NE 43.
4 to 20 mA HART:
• Max. alarm: can be set from 21.5 to 23 mA
• Min. alarm: < 3.6 mA (factory setting)