Liquiphant FTL64 HART
Endress+Hauser
11.3
Diagnostic event in the operating tool
If there is a diagnostic event in the device, the status signal appears on the top left in the
status section of the operating tool together with the associated icon for event behavior
according to NAMUR NE 107:
• Failure (F)
• Function check (C)
• Out of specification (S)
• Maintenance required (M)
Click the status signal to see the detailed status signal.
The diagnostic events and remedial measures can be printed in the Diagnostic list
submenu.
11.4
Adapting the diagnostic information
The event level can be configured:
Navigation: Diagnostics → Diagnostic settings → Configuration
11.5
Queued diagnostic messages
The display alternates between the queued diagnostic messages and the measured value.
Queued diagnostic messages can also be displayed in the Active diagnostics parameter.
Navigation: Diagnostics → Active diagnostics
11.6
Diagnostic list
All the diagnostic messages that are currently queued can be displayed in the Diagnostic
list submenu.
Navigation: Diagnostics → Diagnostic list
Diagnostic
Short text
number
Diagnostic of sensor
004
Sensor defective
007
Sensor defective
042
Sensor corroded
049
Sensor corroded
061
Sensor electronics faulty
062
Sensor connection faulty
081
Sensor initialization faulty 1. Restart device
Diagnostics and troubleshooting
Remedy instructions
1. Restart device
2. Replace electronics
3. Replace device
1. Check fork
2. Replace device
1. Check fork
2. Replace device
1. Check fork
2. Replace device
Replace electronics
1. Check Main to sensor
connection
2. Replace electronics
2. Contact service
Status
Diagnostic
signal
behavior
[from the
[from the
factory]
factory]
F
Alarm
F
Alarm
F
Alarm
1)
M
Warning
F
Alarm
F
Alarm
F
Alarm
43