Page 1
Certificate of Analysis (CoA) are matching and correct. 1. Introduction Thank you for purchasing the Shodex product. Shodex RSpak JJ-50 2D is a multimode (reversed phase and anion exchange modes) column filled with spherical porous particles of polyvinyl alcohol modified with quaternary ammonium group.
Page 2
(3) Recommended sample injection volume is less than 5 μL per column. 7. Column Usage Procedure RSpak JJ-50 2D is a semi-micro column. To achieve its best performance, please use it with a semi-micro type system. 7.1 HPLC System Preparation Wash entire LC system prior to the column installation, including all flow-lines and sample loop by switching the valve, and then replace the washing solution with the eluent to be used.
Page 3
(Refer to section 7.4 Column Cleaning, Method 1). (6) When using a gradient method, changes in the eluent compositions may increase the column backpressure. Adjust the flow rate and column temperature so that the column backpressure remains below the usable maximum pressure. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 4
(a cool and dark space is recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent. Attention! · Never allow inside of the column to dry. It can damage the column. SHOWA DENKO K.K. (https://www.shodex.com/)
Page 5
(2) Do not make a strong impact on the column. Do not drop or hit the column on a hard surface. Refer to the Shodex™ website (https://www.shodex.com/) for product details and their applications. For additional assistance, contact the distributor from whom you purchased the column or contact your regional Shodex...
Page 6
11. Warranty (Ver. 4COLUMN) (1) Showa Denko K. K. warrants that the Shodex product, at the time of delivery to the user, will conform to the specification of the attached CERTIFICATE OF ANALYSIS, if the Shodex product is used in accordance with the attached operating manual.