The Rc:112A Reader - Cirrus Research doseBadge CR:112A User Manual

Reader unit and personal noise dosimeters
Table of Contents

Advertisement

There is no Time History data for
the measurement

The RC:112A Reader

Symptom
The Reader will not switch on
The Reader gives an error when
Resetting the doseBadge:
"No Badge"
The Reader gives an error when
calibrating the doseBadge:
"No Badge"
"Too Low"
The Reader gives an error when
Calibrating the doseBadge:
"Not Compatible"
The Reader gives an error when
downloading (Reading) the
doseBadge:
"No Badge"
The Reader gives an error when
Reading the doseBadge:
"Not Compatible"
There is no Time History data for
the measurement
"Trace Error"
The Reader will not download to
the software
doseBadge User Manual
The Time History setting in the
Reader is configured to Off.
Possible Cause
The Reader Batteries are flat
The doseBadge battery is flat
The InfraRed window on the
doseBadge is not aligned with the
InfraRed Port on the Reader
The ambient light is too high and
is interfering with the Infra Red
Communications.
The doseBadge battery is flat
The InfraRed window on the
doseBadge is not aligned with the
InfraRed Port on the Reader
The doseBadge is not pushed into
the Acoustic Calibrator
The ambient light is too high and
is interfering with the Infra Red
Communications.
The doseBadge is a CR:100A or a
CR:100B and the configuration is
not supported by these units.
The doseBadge battery is flat
The InfraRed window on the
doseBadge is not aligned with the
InfraRed Port on the Reader
The ambient light is too high and
is interfering with the Infra Red
Communications.
The doseBadge is a CR:100A or a
CR:100B and the configuration is
not supported by these units.
The Time History setting is
configured to Off.
There has been an error during
the download of the Time History
Data.
The connection between the
Reader and the PC is not correct
Turn the Time History download
function on
Solution
Replace the batteries in the
Reader unit.
Charge the doseBadge
Align the InfraRed window on the
doseBadge with the InfraRed Port
on the Reader unit
Avoid using the doseBadge in
very bright sunlight. Shield the
Infra Red windows from very
bright light.
Charge the doseBadge
Align the InfraRed window on the
doseBadge with the InfraRed Port
on the Reader unit
Insert the doseBadge into the
Acoustic Calibrator cavity on the
Reader unit
Avoid using the doseBadge in
very bright sunlight. Shield the
Infra Red windows from very
bright light.
Charge the doseBadge
Align the InfraRed window on the
doseBadge with the InfraRed Port
on the Reader unit
Avoid using the doseBadge in
very bright sunlight. Shield the
Infra Red windows from very
bright light.
Turn on the Time History
download option
Ensure that the cable is
connected between the Reader
Page 43