To wake the doseBadge, shake the unit as shown in the diagram below. When the
sensor is activated, the doseBadge status indicator will flash (blue) twice and the
doseBadge can be used.
Note that the sensor may be activated when the doseBadge is removed from the
charger unit or when it is handled. If the indicator light does not flash when the
doseBadge is shaken, check that the doseBadge is not already awake by pointing the
Reader Unit at the doseBadge and pressing the Stop button.
If the indicator shows, the doseBadge is awake and ready to use.
Please note that if the doseBadge is running, it will not go into Sleep mode.
The doseBadge will only enter Sleep mode if it has been stopped for more
than 1 hour and does not receive commands from the Reader unit.
Please note that if the internal battery in the doseBadge is not charged, it will
not be possible to wake the doseBadge from Sleep mode. Ensure that the
doseBadge is fully charged before attempting to use the unit.

Insert the doseBadge in the Acoustic Calibrator

Before the doseBadge can be reset or calibrated, it must be inserted into the Acoustic
Calibrator cavity to allow communication between the doseBadge and the Reader unit.
The figures here show the alignment of the doseBadge in the Reader cavity. Ensure that
the InfraRed window of the doseBadge is aligned with the InfraRed Port of the Reader
unit.