Pifoc long-travel objective scanner, pifoc piezo nanofocusing system for long travel ranges and fast step-and-settle, pifoc high dynamics scanner (40 pages)
Precautionary measures to avoid the risk. INFORMATION Additional information that can affect your application. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 4 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
5. Click the desired manual and fill out the inquiry form. The download link will then be sent to the email address entered. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 5 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
It is only possible to use the P-725 as intended when it is completely mounted and connected. It must be operated with a suitable controller that is available from PI (p. 10). The controller is not included in the scope of delivery of the P-725.
Page 7
Do not use extension cables. If you need longer cables, contact our customer service department (p. 23). Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 7 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
(p. 23). 4. Keep all packaging materials in case the product needs to be returned. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 12 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
You have read and understood the safety precautions (p. 6). The P-725 is not connected to the electronics. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 14 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Note that only the P-725.xxL thread adapter on the objective side is required when mounting the P-725 using the mounting holes in the base body. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 15 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Page 16
5. Screw the objective from below into the thread adapter on the objective side of the P-725. Do not exceed the maximum torque of 0.6 Nm. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 16 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Connect the plug of the P-725 to the corresponding socket on the electronics (refer to the user manual for the electronics). Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 17 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Set the piezo voltage to 0 V on the controller if the P-725 is not used but should remain switched on to ensure the temperature stability. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 18 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
If the connecting cable of the P-725 is accidentally pulled out of the electronics during operation Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 19 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
In open-loop operation: Set the piezo voltage to 0 V on the electronics. Discharging a P-725 not Connected to the Electronics Connect the P-725's voltage plug to the switched-off electronics from PI for 10 seconds. Maintenance NOTICE Misalignment from loosening screws! The P-725 is maintenance-free and precision adjusted.
Screw the objective out before removing. Hold the P-725 firmly when you loosen the upper thread adapter. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 21 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Page 22
4. Use the wide side of the key tool to unscrew the thread adapter on the microscope side. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 22 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
User Manual P725UM0001EN, applies to P-725.xCDE2 CBo/KSch, 4/1/2022 Customer Service For inquiries and orders, contact your PI sales representative or send us an email (service@pi.de). If you have questions concerning your system, provide the following information: − Product and serial numbers of all products in the system −...
Page 24
10 to 70 °C 10 to 70 °C The resolution of the system is limited only by the noise of the amplifier and the measuring technology because PI piezo nanopositioning systems are free of friction. All specifications based on room temperature (22 °C ±3 °C).
* The specifications apply to room temperature (22 °C ± 3 °C). The specifications can deviate outside of this range. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 25 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
P-725.1CDE2 and P-725.4CDE2, dimensions in mm (please order adapters and spacers separately). Note that a comma is used in the drawings instead of a decimal point. Physik Instrumente (PI) GmbH & Co. KG, Auf der Römerstraße 1, 76228 Karlsruhe, Deutschland Page 26 / 28 Phone +49 721 4846-0, Fax +49 721 4846-1019, Email info@pi.de,...
Dispose of your old equipment according to international, national, and local rules and regulations. To fulfill the responsibility as the product manufacturer, Physik Instrumente (PI) GmbH & Co. KG undertakes environmentally correct disposal of all old PI equipment made available on the market after 13 August 2005 without charge.