3 - 6
Checking the ground wire connections of the ON/OFF assembly
Check the ground wire connections prior to startup.
To do this, plug the monitor trolley cable coming from the monitor trolley into the C-arm
chassis (X10).
Measure the ground wire resistance from the ground wire contact of the power plug to the
following measuring points:
from
Ground wire contact
on the power plug
of the power cable
Tab. 1
Connecting the UPS
Connect the power input plug and power output plug to the UPS.
ARCADIS
System Manual
Installing the ON/OFF assembly
to
Monitor trolley, ON/OFF assembly
ground wire contact
power cable (1/Fig. 2)
Monitor trolley, ON/OFF assembly
ground wire contact
(5/Fig. 2)
Monitor trolley, side cover,
ground wire connection
Monitor trolley, PC (image processor),
ground wire connection
Monitor trolley, live monitor,
ground wire connection
Monitor trolley, reference monitor,
ground wire connection
Monitor trolley, optional printer (if present),
ground wire connection
C-arm chassis, contact for
equipotential bonding
C-arm chassis image intensifier housing
n.a.
SP00-000.841.01
Rev. 01
Page 6 of 10
08.05
CS PS 24
3
Limit value:
< 0.2 ohms
3
Siemens AG
Medical Solutions