Inspection and maintenance
3.2.9

Power cable

NOTE
PMP
Cable winch replacement
If the replacement interval has been reached, replace the cable winch according to the
"Replacement of parts" instructions.
SIM
Checking the power cable
Pull the entire power cable (6m) out of the cable winch. Check the cable for damage.
SIM
General cable winch information
Pull the entire power cable (6 m) out of the cable winch. Hold the cable while coiling and
move it up and down until the cable is fully coiled. Verify that the cable winch and the
brake work properly.
PMP
Cleaning
Clean the power cable with a cloth. Dampen the cloth with a mixture of lukewarm water
and household cleaner.
3.2.9.1
Power plug
SIM
Checking the power plug
Check the power plug for proper functioning.
Open the plug and check the terminals and strain relief.
Fig. 14: Ground wire
3.2.10

Lubrication

PMP
Pull-bar slide bushing
Siemens AG
Medical Solutions
Replacement of cable winch including cable:
MOBILETT XP Digital every 4 years
The ground wire has to be slightly longer than the other wires (applies to all plug
types).