Do you have a question about the BSC-3FA2-NA and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Biobase BSC-3FA2-NA

  • Page 1 Biological Safety Cabinet Class II A2 BSC-3FA2-NA/ GL User Manual BIOBASE GROUP Version 2021.06...
  • Page 2: Preface

    Preface Thank you very much for purchasing our class II A2 biological safety cabinet. Please read the “Operating Instructions” and “Warranty” before operating this unit to assure proper operation. After reading these documents, be sure to store them securely together with the “Warranty” at a hand place for future reference.
  • Page 3: Table Of Contents

    Content Preface..............................1 Content..............................2 1.Unpacking, Installation, Debugging....................4 1.1 Unpacking..........................4 1.2 Accessories checking........................ 6 1.3 Installation conditions and using environment................7 1.4 Installation..........................7 1.5 Checking after installation.......................11 1.6 On-site debugging........................12 1.7 Test and standard........................12 2. User Instructions..........................14 2.1 Product instruction........................14 2.1.1 Product concept......................
  • Page 4 2.5.2 Structural material maintenance..................22 2.5.3 Clean process.........................22 2.5.4 Vent tank........................22 2.6 Replacement part list....................... 23 2.7 Wiring diagram........................24 3.Common faults & solution......................25 3.1 Common faults & solution...................... 25 3.1.1 Warning and reminder....................25 3.1.2 Trouble shooting......................25 3.1.3 Simple accessories replacement..................27 3.2 Label Description........................
  • Page 5: Unpacking, Installation, Debugging

    1.Unpacking, Installation, Debugging Please firstly check if packing box is in good condition. If the packing box is damaged, please take photos. 1.1 Unpacking Choose the proper unpacking method according to the actual situation. Method 1 Necessary tools for unpacking: Electric drill with hexagon dead M8 Picture 1 Method 2 Use M8 Wrench to unpack Picture 2...
  • Page 6 Picture 3 Notes: 1) Check the package after receiving the device. Broken glass or other visible damage must be indicated in the equipment receipt, and immediately notify the shipper. Note: Equipment cannot be returned to the factory, unless receive return authorization! 2) If the device is transported in cold weather, please place it in the area for 24 hours at room temperature before putting it in working place.
  • Page 7: Accessories Checking

    1.2 Accessories checking Refer to the packing list and check the accessories and materials BSC-3FA2-NA/ BSC-3FA2-GL Packing list Items Model Position Quantity Main body Inside the packing box 1 unit On top or at the back of main UV lamp (T6 18W)
  • Page 8: Installation Conditions And Using Environment

    1.3 Installation conditions and using environment To avoid disturbances to the safety cabinet and its operator, follow the following guidelines, while determining a suitable location for the cabinet: a. The distance from the plane of the aperture to any circulation space should be at least 1000 mm, so as to preserve a zone undisturbed by anyone other than the operator.
  • Page 9 components. If necessary, test filter and the fixed bolt on the front panel, these bolts should not be too tight. If you need to check the surface of the air filter, you can loosen bolts fixed well wind network and take it out carefully from the front. 7) Check if the work place stable.
  • Page 10 Picture 5 Aligned the mounting holes on the bottom of the cabinet side with the locating bolts, slow down the cabinet on the mounting base. Picture 6...
  • Page 11 Take out the M10*20 Hexagon socket head screws, Flat washer 10, Spring washer 10 from the accessory box, and fasten tightly according to Picture 6. c. Installation of Drain valve Picture 7 1. Drain valve connector 2. Shim 3. Safety cabinet bottom installation holes 4.
  • Page 12: Checking After Installation

    Picture 9 1. Tighten nut 2.The tap/gas tap Pick up the water tap, gas tap and tightening nut, install as shown in Picture 9. 1.5 Checking after installation Checking Items Normal situation Wind speed display Inflow100-110fpm, downflow 60-70 fpm Pressure display exhaust filter80-120Pa, downflow filter 80-120Pa Fan operation The fan running...
  • Page 13: On-Site Debugging

    3) Press the LIGHT button on control panel, turn on the lights. If the lamp is not bright, please check lamp-tube or socket, it may course loose between the lamp and lamp holder in the carrying process. 4) Press UV button on the control panel; turn on the UV lamp and to check whether UV light is working properly.
  • Page 14 test should be conformed the rules of the NSF49. The opening air velocity testing data see the APPENDIX 1. To adjust the fan speed adjustment should be according to the data specified in the APPENDIX 1. To set the air velocity of the opening work area correctly. We emphasis again that only the qualified people can use the calibrated equipments to carry out the adjustments.
  • Page 15: User Instructions

    2. User Instructions 2.1 Product instruction 2.1.1 Product concept Biological safety cabinets meet the standards of NSF/ANSI49-2012. However, the A/B3 is the old version. The equipment has the following three functions: Protect the operators against the hazardous particles or hazardous particles in the air. Provide the clean working area, prevent the air pollutants from touching the working bench.
  • Page 16: Circuit Of Digital Control

    (f) When the cabinet is running, it can storage clean instruments, but not too much. (g) When sampling operation is finished, the samples should be moved to the left or right of the sampling dish. Correct operation can make sure there is no cross infection of the samples in the cabinet.
  • Page 17: Using Of Uv Lamp

    Whether the safety cabinet is working, the exhaust hood can always provide continuous stable gas flow exhausting outdoor. The total exhaust air rate of the exhaust hood is no less than 100m3/h of safety cabinet’s gas displacement, gas displacement of safety cabinet BSC-3FA2-NA/ BSC-3FA2-GL is separately 420-465m3/h.
  • Page 18: 11Technical Parameter

    3)Flow alarm system instruction Flow alarm system is an independent software package programmed in advance to monitor the abnormal of air volume. It alarms when the air volume exceed normal value. 2.1.11Technical Parameter Model BSC-3FA2-NA BSC-3FA2-GL Technical parameter Rated voltage AC 110V±10% □...
  • Page 19: Structural Composition

    2.2 Structural composition Picture 12...
  • Page 20: Control Panel

    1.Air Outlet safety guard 5.Alarm rest plate 2.Control panel 6.UV lamp 3.LCD display 7. IV bar 4.Fluorescent lamp 8. Base stand 2.3 Control panel Picture 13 1.UV lamp 13.Glass window up status 2.Fluorescent lamp 14. UV lamp working time 3.Blower 15.
  • Page 21: The Usage Of Remote Control

    : The power button: Master switch to control other function keys : To control fluorescent lamp. When you press each time, the status of the fluorescent lamp and the corresponding status counter on the LCD change at a time from bright into extinguish or the opposite.
  • Page 22 Functions of Control panel: 1.Power 2.Sub 3.Install Timer 4.Confirm 5.Cancel 6.Turn Up(+) 7.Turn Down(-) 8.Blower status(FAN) 9.UV lamp(UV) 10.Fluorescent lamp (LIGHT) Picture 14 11.Socket status(SOCKET) 12.Mute status (MUTE) 13.Glass window up status(UP) 14.Glass window up status(DOWN) Remote Control A. Reservation Time (SUB) a.
  • Page 23: Methods And Procedures For Disinfection

    Biological safety cabinet is equipped with special UV lamp. When turning on or turning off the cabinet, sterilization time of UV lamp should be at least 30 minutes. In order to save the waiting time of turning on or turning off the cabinet, we develop reservation time function. It realizes function of automatic turning on or turning off the cabinet after the sterilization finished.
  • Page 24: Replacement Part List

    -Before working -Before changing the filter -Before performance checking(unless it is a new machine and never used) -Before moving the equipment -Start an entirely different working program -When the safety office or safety committee requires Note: Only the safety office or safety committee can determine the right disinfection method. 2.6 Replacement part list Designation Specification...
  • Page 25: Wiring Diagram

    2.7 Wiring diagram Picture 15...
  • Page 26: 3.Common Faults & Solution

    3.Common faults & solution 3.1 Common faults & solution 3.1.1 Warning and reminder Digital display of pressure difference, digital velocity display, audible and visual alarm system. 1) Over safety height alarm for front window. There will be audio and visual alarm when front window is lifting over safety height. Same time LCD display will twinkle exclamation mark.
  • Page 27 Faults Measures (a)Check whether the power socket inserted tightly (b)Check whether these three insurance tube of circuit board burned out :zero Equipment line insurance tube F3 12.5A, fire line insurance tube F2 12.5A, transformer doesn’t work insurance tube F1 125 mA. (c)Transformer fault or cable poor contact can lead to circuit board working voltage abnormal.
  • Page 28: Simple Accessories Replacement

    indicate the model and serial number of the cabinet purchased. 3.1.3 Simple accessories replacement 1) Replace fluorescent light When replacing lights, make sure that the power is off, open the operation panel like shown in Picture 16, use the control panel support frame (fixed in the inside position of the control panel as shown), then as Picture 17 shown rotary screw off the tubes, take the correspondence type of lamp, put it to the lamp holder and rotary screw to the right position in the opposite direction.
  • Page 29: Label Description

    UV lamp Picture 18 3.2 Label Description 1)Biological hazard label Picture 19 2)Ground symbol Picture 20 3)Glass door super elevation warning label Picture 21 4)Warning note Picture 22...
  • Page 30: Warranty

    5)DECONTAMINATE note Picture 23 4. Warranty We provide warranty for all Biosafety Cabinet that purchase from our company,and will abide by the following terms and conditions: (a) 18 months warranty period is as follows; Warranty is 12 months after delivery,include the machine and parts; The following 6 months only guarantee parts;...
  • Page 31 APPENDIX 1 Biosafety Cabinet The setting of the downflow and inflow velocity Measured by Air Flow Anemometer Model type Inflow Velocity Downflow Velocity 100-110fpm 60-70fpm BSC-3FA2-NA/ BSC-3FA2-GL 0.53±0.025m/s 0.33±0.025m/s...
  • Page 32 BIOBASE GROUP 2# building, No.9 Gangxing Road, High-tech Zone, Jinan City, Shandong Province, China Tel: +86-531-81219803/01 Fax: +86-531-81219804 Inquiry: export@biobase.com Complaints: customer_support@biobase.cc After-sales service: service_sd@biobase.cc; service_ivd@biobase.cc Web: www.biobase.cc/www.meihuatrade.com / www.biobase.com...