ELGA CENTRA S200 US Operator's Manual page 29

Table of Contents

Advertisement

ELGA
Alarm settings QS1 screen
Alarm settings QS2 screen
Alarm settings TS1 screen
Alarm settings TS2 screen
Step 9 - RO permeate quality alarm settings QS1
Step 10 - Product water purity alarm settings QS2
Step 11 - RO permeate temperature alarm settings TS1
Step 12 - Product water temperature alarm settings TS2
CENTRA - S200 US/CENTRA - R200 US Version 01 – 08/14
CENTRA - S200 US/CENTRA - R200 US Operator Manual
This screen is used for setting the value at which the RO
permeate quality alarm will activate. This alarm does not
stop RO production but can extend or instigate flush
routines.
1.
PRESS
SCROLL
(Increments of 10, ranging from 20 to 100µS/cm).
If '0' is selected the alarm will be tuned 'OFF'.
PRESS TICK  to proceed to step 10.
PRESS TICK .
2.
This screen is used for selecting the value at which the
product water purity alarm will activate. The alarm does
not stop the process and will automatically reset if the
purity level recovers.
1.
PRESS SCROLL
(increments of 1, ranging from 1 to 17MΩ.cm -
CENTRA - R200 US
increments of 10, ranging from 20 to 100µS/cm).
Note:
The product water alarm points are only
displayed in M
regardless of the water purity unit displayed
setting.
TICK  to proceed to step 11.
PRESS TICK .
2.
This screen is used for selecting the value at which the
RO permeate temperature alarm will activate. The alarm
does not stop the process and will automatically reset if
the temperature falls below the set point.
1.
PRESS
SCROLL
(increments of 1, ranging from 20 to 50°C)
PRESS TICK  to proceed to step 12.
PRESS TICK .
2.
This screen is used for selecting the value at which the
product water temperature alarm will activate. The alarm
does not stop the process and will automatically reset if
the temperature falls below the set point.
1.
PRESS
SCROLL
(increments of 1, ranging from 20 to 50°C)
PRESS TICK  to proceed to step 13.
PRESS TICK .
2.
Note:
In addition to the alarm, if the water temperature
rises above 55
alarm, stop and await operator intervention -
to
select
alarm
OR
to select alarm point
OR
.cm and do not change to µS/cm
OR
to
select
alarm
OR
to
select
alarm
OR
°
C in the system the unit will
point
point
point
Page 23

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Centra r200 usCentra-s200hfv usCentra-r200hfv us

Table of Contents