Calculation for DI Cartridge (LC207) Cartridge life ... 28 TROUBLE SHOOTING ............29 KEY TO CONTROL PANEL ..........30 TECHNICAL SPECIFICATION ........... 31 WARRANTY/CONDITIONS OF SALE ....... 33 USEFUL CONTACT DETAILS ........... 35 MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 2...
Electricity It is essential that the electrical supply to the MEDICA Pro EDI is isolated before any items are changed or maintenance work performed. The appliance coupler (mains lead) connected to the rear of the unit can be removed to isolate the power supply. If access to this is restricted then it is recommended that access to the supply socket is easily available to disconnect the electrical supply.
Control of Substances Hazardous to Health LC136 (COSHH) LC105 LC177 LC207 Material safety data sheets covering the various replaceable components are available upon request. Contact your local ELGA LabWater distributor. Consumables. Consumables Typical Max. Shelf Service Life* Life...
Installation The unit is supplied with a quick start manual that allows the unit to be quickly installed and commissioned. It is recommended that a trained ELGA representative carry out the installation of the product. Commissioning The unit is supplied with the software set in a commissioning mode that must be completed before being used to supply the application.
MEDICA Pro EDI Operator Manual ELGA CONTROLS The MEDICA Pro EDI control panel has a range of control icons. General icons are as follows. Further icons are described in the appropriate sections a complete listing is included in Section 9.
ELGA Service Engineers. System Preferences When the MEDICA Pro EDI unit is started for the first time after completing the commissioning routine the following steps should be carried out to set up your choices: Note: Your choices can be changed and implemented during normal operation.
Page 10
PRESS to change mode ( = ON) PRESS ✔ to proceed to step 5. PRESS ✔ . Audible alarm enable screen Note: The visual alarm cannot be disabled. MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 8...
Page 11
ON) PRESS ✔ to proceed to step 9. PRESS ✔ . Note: This mode is for validation purposes only and should not be selected as standard. Uncompensated water quality screen MEDICA Pro EDI MANU39885 Ver 1 – 05/12 Page 9...
Page 12
PRESS to select alarm point (increments of 1, ranging from 20 to 50C) PRESS ✔ to proceed to step 12. PRESS ✔ . Alarm settings TS1 screen MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 10...
Page 13
30 minutes. PRESS to enter times. Setting period of operation PRESS ✔ . Note: Night service will only be operational if Continuous Operation is not enabled. MEDICA Pro EDI MANU39885 Ver 1 – 05/12 Page 11...
Page 14
MEDICA Pro EDI Operator Manual ELGA Step 15 - Operational days selection (screen 031) Select the days that the MEDICA Pro EDI is required to operate by highlighting the relevant box. Monday = 1, Sunday = 7 PRESS to enter selection screen PRESS ✔...
Page 15
Feed Water Quality Screen PRESS ✔ to accept Step 19 - Data output (screen 035) The MEDICA Pro EDI can transmit operational data to an Xport storage device. PRESS to format the storage device.
Page 16
ELGA Step 21 - Programming of User Keys (screen 037) During the life of the MEDICA Pro EDI you may need to delete or add User keys to prevent or allow access to user choices. This feature is only available to the Master key holder.
PRESS ✔ to accept replacement date and proceed to Protek Filter Replacement. PRESS ✔ confirm that resetting is required PRESS ✗ to abort reset. PRESS ✔ . UV replacement date screen MEDICA Pro EDI MANU39885 Ver 1 – 05/12 Page 15...
Page 18
Sanitization replacement date PRESS ✔ to accept replacement date and complete settings. PRESS ✔ confirm that resetting is required PRESS ✗ to abort reset. PRESS ✔ . Sanitization reminder screen MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 16...
Night service/Operational day (screen 030) Refer to Section 3.2 - Step 14 and 15. Your MEDICA Pro EDI can be programmed to operate on specific days between selected times. This is to optimize the efficiency of the unit and to minimize rises in water temperature.
WRITE the installation date on the label of the filter for future reference. INSTALL filter. CVF position SWITCH on power. RESET consumable reminder as described in Section 3.3, Step 3. General View MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 18...
EASE back and ENSURE that the pack is fully engaged (down) in the retainers. SWITCH on power. FOLLOW screen prompts to accept new cartridge replacement date. RESET reminder as described in section 3.3, Step 2. MEDICA Pro EDI MANU39885 Ver 1 – 05/12 Page 19...
4. UNSCREW UMF from top connection 5. INSTALL new UMF 6. RECONNECT 7. RINSE water to drain to recommission the system. Ultra-Microfilter 8. RESET reminder as described in section 3.3, Step 6. Ultra-Microfilter position MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 20...
RECONNECT to electrical connectors top and bottom. REFIT retaining spring clips top and bottom REPLACE Medpure pack (if fitted). RESET consumable reminder as described in Section 3.3, Step 4. UV Lamp connection MEDICA Pro EDI MANU39885 Ver 1 – 05/12 Page 21...
11. CLOSE retaining clamp and fasten CLIP Protek L2 (LC177) 12. The unit will automatically recognize the new cartridge and date. 13. RESET reminder as described in section 3.3, Step 5. Protek L2 (LC177) MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 22...
PRESS ✓ to enter RO sanitization. Enter RO sanitization UNSCREW the large white lid on the left of the unit. ADD 2 CT3 tablets. REPLACE lid. Large white cap removal MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 24...
Once the step within the sanitization is complete the unit will go straight into process and continue to fill the recirculation reservoir RO sanitization process steps PRESS ✓ to accept the sanitization reminder reminder Set Sanitization MEDICA Pro EDI MANU39885 Ver 1 – 05/12 Page 25...
Wash out the mouth thoroughly with water and give large quantities of milk to drink. Obtain MEDICAL ADVICE. Note: A comprehensive Data Sheet is available on request and is supplied with each packet of tablets MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 26...
Remove yellow plastic cover 3 Open valve 3 Close valve 3 when there is no requirement for purified water. Monitor water purity independently or install new LC207 cartridges to guarantee water purity. MEDICA Pro EDI MANU39885 Ver 1 – 05/12 Page 27...
2 hr if the analyzer consumes 20 l/hr. If flows start to decline this may be improved by changing the UMF. The capacity will be less (typically 80% less) if a water purity of 10 M.cm is needed. MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 28...
The appropriate action to then take (for the more common alarm displays) is shown on the Quick Reference Guide (fitted to the inside of the door of the MEDICA Pro EDI.) If the unit cannot be repaired using the information shown, please call your local ELGA LabWater representative (See Section 12.0 - Useful Contact Details).
Calibration Option ON point Cancel Output Clock Recirculate Date Reset Right Sanitization Down PASSkey Sanitization Drain reminder Save data Scroll Transport Hazard mode Feed Fill Viewing angle Add chemical Recirc MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 30...
Technical Support at ELGA LabWater. Pre-treatment LC177 and LA727 (LC123) included in starter kits LC238 and LC237 respectively. MEDICA Pro EDI MANU39885 Ver 1 – 05/12 Page 31...
Page 34
Bacteria <1 CFU/ml** Particles 0.05 µm System to be regularly sanitized and installed following ELGA LabWater installation design guidelines. As part of our policy of continual improvement, we reserve the right to alter the specifications given in this document MEDICA Pro EDI MANU38995 Ver 1 – 05/12...
ELGA MEDICA Pro EDI Operator Manual WARRANTY/CONDITIONS OF SALE ELGA LabWater is a trading name of VWS (UK) Ltd. General Limited Warranty VWS (UK) Ltd. warrants the products manufactured by it against defects in materials and workmanship when used in accordance with applicable instructions for a period of one year from the date of shipment for the products.
Page 36
VWS (UK) Ltd. warrants its products against defects in materials and workmanship as described in the Warranty statement on the preceding pages. MEDICA Pro EDI MANU38995 Ver 1 – 05/12 Page 34...
Tel: +44 (0) 1494 887500 Fax: +44 (0) 1494 887505 E-mail: techsupport@elgalabwater.com For the address of the nearest ELGA LabWater Sales and Service office visit the country list on our website. http//www.elgalabwater.com or contact ELGA at the number above. ELGA is the global laboratory water brand name of Veolia Water.