Table of Contents
loading
Need help?

Need help?

Do you have a question about the PURELAB Option-Q 7/15 - US and is the answer not in the manual?

Questions and answers

Summary of Contents for ELGA PURELAB Option-Q 7/15 - US

  • Page 1 PURELAB Option-Q 7/15 - US Operator Manual...
  • Page 3 © VWS (UK) Ltd. 2014 All rights reserved. Publication ref: MANU40551 Version 3 02/16 ELGA LabWater is a trading name of VWS (UK) Ltd. ELGA® and PURELAB® are registered trademarks. Minncare is a trademark of Minntech Corporation. EfferSan™ is a trademark of Activon Inc.
  • Page 4: Table Of Contents

    Replacement of LC143 Reverse Osmosis Cartridge(s) ............29 SANITIZATION PROCEDURE ........30 Liquid sanitization ..........30 Tablet sanitization..........34 TROUBLE SHOOTING ..........39 CONSUMABLES AND ACCESSORIES ..... 40 KEY TO CONTROL PANEL ........41 Page ii PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 5 11.1 Icons ..............41 11.2 Alarm Conditions ..........41 11.3 Replacement Timers .......... 41 11.4 Low Level, Quality and Standby Alarms ..... 42 WARRANTY/CONDITIONS OF SALE ......43 USEFUL CONTACT DETAILS ........45 PURELAB Option-Q 7/15 - US Version 3 02/16 Page iii...
  • Page 7: Introduction

    1.3 Customer Support PURELAB Option-Q Service support and consumable items are available from your local supplier or distributor. Refer to customer service contact details shown at the end of this publication. PURELAB Option-Q 7/15 - US Version 3 02/16 Page 1...
  • Page 8: Health And Safety Notes

    UV light. WARNING! UNDER NO CIRCUMSTANCES SHOULD LAMP CONNECTED ACTIVATED WHEN OUTSIDE HOUSING. UV Lamp Page 2 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 9: Sanitization Chemicals

    1% of Minncare Cold Sterilant has a pH of 3.5. 2.5 Control of Substances Hazardous to Health (COSHH) Material safety data sheets covering the various replaceable cartridges are available upon request. Contact your local supplier or distributor. PURELAB Option-Q 7/15 - US Version 3 02/16 Page 3...
  • Page 10: Product And Process Description

    Accessories, for detail). Dispense tap Sanitization port Power switch ON/OFF Mains power socket Removable cover Control Panel Fuse Alternative dispense Feedwater inlet tap position connection Door Removable cover PURELAB Option-Q Page 4 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 11: Process Description

    0.2m bacterial filter can be fitted to the dispense tap for added protection.  During periods of non-use the unit will automatically operate in intermittent re-circulation mode to maintain water purity with maximum efficiency. PURELAB Option-Q 7/15 - US Version 3 02/16 Page 5...
  • Page 12 Purified water inlet Inlet outlet Inlet Temp sensor Sanitization Port Flow Strainer controller Ion-exchange Cartridge pack Solenoid LC163 Pre-treatment LC140 LC118 Concentrate Drain Boost pump (optional) Process Flow - PURELAB Option-Q Page 6 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 13: Technical Specifications

    Wall, bench or under bench mounted. mounted. Environment Clean dry indoor. Clean dry indoor. Temp 5 - 40°C. Temp 5 - 40°C. Humidity max 80% Humidity max 80% non-condensing. non-condensing. PURELAB Option-Q 7/15 - US Version 3 02/16 Page 7...
  • Page 14 RS232 Remote display connection Volt free contact-internal Safety Features Power fail safe Boost pump protection from particulates Re-circulation pump protection from particulates Low operating voltage 24V Audio visual alarms Adjustable alarm settings Page 8 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 15 Subject to correct operating and maintenance procedures. As part of our policy of continual improvement we reserve the right to alter the specifications given in this document. PURELAB Option-Q 7/15 - US Version 3 02/16 Page 9...
  • Page 16 PURELAB Option-Q Reverse Osmosis Capacity Charts Option-Q7 Pressure (bar) Graph 1 - Nominal Flowrate vs Inlet Pressure for PURELAB Option-Q7 Option-Q15 Pressure (bar) Graph 2 - Nominal Flowrate vs Inlet Pressure for PURELAB Option-Q15 Page 10 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 17: Controls

    The PURELAB Option-Q control panel has a range of control icons control button control button as follows: Control Panel Button Icon Description LEFT Menu Scroll RIGHT Reset Mute Alarm Accept Printer PURELAB Option-Q 7/15 - US Version 3 02/16 Page 11...
  • Page 18: Installation Instructions

     Maximum Relative Humidity 80% @ 31°C decreasing linearly to 50% @ 40°C, non-condensating The unit is in Installation Category II, Pollution Degree 2, as per IEC1010-1. Unit Rear Mounting Points Page 12 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 19 Feed Water Strainer PURELAB Option-Q RESERVOIR Level control cable Pump Inlet feed All tube connections 8mm (5/16") Drain PURELAB Option-Q Unit Installed with Storage Reservoir PURELAB Option-Q 7/15 - US Version 3 02/16 Page 13...
  • Page 20: Connecting Up The Purelab Option-Q

    Pull plug Depress collet Clean square cut end Tube 8mm OD Feedwater Inlet Connections Feedwater Inlet Connections Simply push in tube to attach Tube in secured position Fitting Tubes Page 14 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 21 Step 3 - Connect High/Low Level Switch to Reservoir 1. INSERT jack plug into the level control socket located at rear of unit and reservoir. Level control Reservoir Level Connections PURELAB Option-Q 7/15 - US Version 3 02/16 Page 15...
  • Page 22 13. Switch water supply on. side panel 14. Press the process button. 15. Ensure correct operation of dispense tap. Manifold Panel screws fixing screws Replacing side panel Positioning the Dispense Tap Page 16 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 23: Initial Controller Set-Up

    Select the  option required using the button and accept with the button. Audible Alarm Enable/Disable PURELAB Option-Q 7/15 - US Version 3 02/16 Page 17...
  • Page 24  Press to reset timer or press to abort reset.  Press to jump to the next consumable. Ion-exchange Cartridge Timer Page 18 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 25  UV tube replacement timer (hours remaining).  Pre-treatment cartridge replacement (hours remaining). Step 5 - Report Printing (Only if connected to LA618) To print a report press the icon. Water Conductivity PURELAB Option-Q 7/15 - US Version 3 02/16 Page 19...
  • Page 26: Initial Start Up

    13. REMOVE the by-pass block. (Store in the front door). Inserting 14. INSERT Ion-exchange cartridge LC163. (See Section 7.2). Removing & Inserting By-pass Block 15. PRESS the PROCESS button to begin the water purification process. Page 20 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 27: Operation

    1 hour. The unit will re-circulate the reservoir contents for 10 minutes every hour. Pressing the PROCESS button at any time during this mode will initiate re-circulation. Flash Normal Process Screens PURELAB Option-Q 7/15 - US Version 3 02/16 Page 21...
  • Page 28: Alarm Conditions

    Press the button to mute the alarm. If water purity stays outside acceptable limits replace the ion-exchange pack following the instructions in Section 7.2. Flash Water Purity Alarm / Standby Page 22 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 29 This alarm condition will cause the process to turn off. Connect the Reservoir Level reservoir level and press the process key to clear the alarm. Disconnect Alarm PURELAB Option-Q 7/15 - US Version 3 02/16 Page 23...
  • Page 30: Maintenance

    Standard conditions are 4bar (60psi) inlet pressure at 15°C, potable water with clean pre-filter. CAUTION! Ensure that the display and replacement timer settings are reset after replacing consumables. (Refer to Section 5.4). Page 24 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 31: Replacing The Lc140 Pre-Treatment Cartridge

    5. RESET pre-treatment cartridge timer. (See Section Setting Display Replacement Timer). 6. PRESS the PROCESS button to start the unit. 7. CHECK the unit for leaks and close front door. PURELAB Option-Q 7/15 - US Version 3 02/16 Page 25...
  • Page 32: Replacing The Lc163 Ion-Exchange Cartridge Pack

    7. RESET ion-exchange pack timer. (See Section 5.4 - Setting up Display and Replacement Timer). Replacing Ion-exchange 8. PRESS the process button to start the unit. Cartridge 9. CHECK the unit for leaks. Page 26 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 33: Replacing The Ultraviolet Lamp (Lc118)

    Lamp housing. 2. SLIDE the new UV lamp into the center bore of Stainless steel the UV housing. chamber 3. Note orientation of pins on each end. LC118 Lamp PURELAB Option-Q 7/15 - US Version 3 02/16 Page 27...
  • Page 34 5. REFIT the UV housing on securing screws. 6. TIGHTEN securing screws. Step 8 - Set UV Change Reminder 1. SEE Section 5.4 - Setting Up Display and Replacement Timers. Page 28 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 35: Cleaning The Re-Circulation Strainer

    Customer Service. WARNING! ALL NEW RO CARTRIDGES ARE FILLED WITH A BACTERIOSTATIC SOLUTION TO PREVENT BACTERIAL CONTAMINATION DURING STORAGE. THE RO CARTRIDGES WILL THEREFORE REQUIRE RINSING PRIOR TO USE. PURELAB Option-Q 7/15 - US Version 3 02/16 Page 29...
  • Page 36: Sanitization Procedure

    The sanitization frequency required is dependent on the feedwater, local environment, usage patterns and application. As a general rule, ELGA LabWater recommends sanitization of the RO at 6 monthly intervals. However, the period between sanitizations could be extended in particular circumstances. For example, microbial growth will usually be lower with clean feed water, well-maintained pre- treatment, low temperatures and heavy usage.
  • Page 37 Step 5 - Fit By-pass Block 1. WET ‘O’ rings on by-pass block. 2. SLIDE by-pass block into unit. 3. ENSURE by-pass block is locked in place. Inserting Fit Sanitization By-pass Block PURELAB Option-Q 7/15 - US Version 3 02/16 Page 31...
  • Page 38 3. PLACE in storage area. Step 10 - Replace Ion-exchange Cartridge Pack 1. INSERT ion-exchange cartridge into Removing compartment. By-Pass Block 2. ENSURE the cartridge is locked in position. 3. CLOSE front door. Page 32 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 39 ELGA LabWater technical support or your local representative should be contacted for advice and assistance. It is recommended that only ELGA LabWater service personnel or other fully trained staff should perform the sanitization procedure for the docking vessel/reservoir...
  • Page 40: Tablet Sanitization

    3. UNSCREW cap on sanitization by-pass block. 4. INSERT ONE EfferSan™ tablet. By-pass Block 5. REFIT cap on by-pass block, hand tight. Page 34 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 41 Step 9 - Return to Normal Operation 6. TURN on power. 7. RECONNECT the outlet tube to the reservoir. 8. PRESS PROCESS button to return to normal operation. 9. CHECK the system for leaks. PURELAB Option-Q 7/15 - US Version 3 02/16 Page 35...
  • Page 42 2. PRESS the PROCESS button to start the sanitization process. 3. ALLOW the sanitization PROCESS to continue for 60 minutes. 4. PRESS the PROCESS button to stop the sanitization process. Page 36 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 43 Replace Ion Exchange Cartridge 1. TURN on power. 2. RECONNECT the outlet tube to the reservoir. 3. PRESS PROCESS button to return to normal operation. 4. CHECK the system for leaks. PURELAB Option-Q 7/15 - US Version 3 02/16 Page 37...
  • Page 44 Mouth - wash out the mouth thoroughly with water and give large quantity of milk to drink. Obtain medical advice. Note: A comprehensive Data Sheet is available on request and is supplied with each packet of tablets. Page 38 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 45: Trouble Shooting

    Replace UV Lamp. (See Section 7.3 - Replacing the Ultraviolet Lamp). Unit noisy Open front door and secure pipework to stop vibration. No flow from dispense Push PROCESS button. Check reservoir level is >40%. PURELAB Option-Q 7/15 - US Version 3 02/16 Page 39...
  • Page 46: Consumables And Accessories

    LA605 RS232 Printer kit LA618 RS 232 Remote display kit LA675 Remote dispense valve LA521 Chlorine test kit TEST30679 Docking Vessel – DV25 LA621 Pre-filter if boost pump fitted LA582 Page 40 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 47: Key To Control Panel

    11.3 Replacement Timers Replacement Timer Icon Preset Ion-exchange Cartridge 4380 hours ( = 6 months) UV Lamp 8760 hours ( = 12 months) Pre-treatment Cartridge 4380 hours ( = 6 months) PURELAB Option-Q 7/15 - US Version 3 02/16 Page 41...
  • Page 48: Low Level, Quality And Standby Alarms

    PURELAB Option-Q 7/15 - US Operator Manual ELGA 11.4 Low Level, Quality and Standby Alarms Screen Description Low level alarm Process on and water purity alarm Standby and purity alarm Standby position Page 42 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 49: Warranty/Conditions Of Sale

    ELGA PURELAB Option-Q 7/15 - US Operator Manual 12. WARRANTY/CONDITIONS OF SALE ELGA LabWater is a trading name of VWS (UK) Ltd. General Limited Warranty VWS (UK) Ltd. warrants the products manufactured by it against defects in materials and workmanship when used in accordance with applicable instructions for a period of one year from the date of shipment for the products.
  • Page 50 VWS (UK) Ltd. warrants its products against defects in materials and workmanship as described in the Warranty statement on the preceding pages. Page 44 PURELAB Option-Q 7/15 - US Version 3 02/16...
  • Page 51: Useful Contact Details

    +44 (0) 203 567 7300 Fax: +44 (0) 203 567 7205 E-mail: techsupport@elgalabwater.com For the address of your nearest ELGA LabWater Sales and Service office visit the country list on our website http://www.elgalabwater.com or contact ELGA at the number above.