Summary of Contents for ELGA PURELAB 7000 - US

  • Page 1 PURELAB 7000 - US Quick Start Manual...
  • Page 3: Table Of Contents

    5.1.1 RO + Recirculation loop sanitization ......8 Tablet sanitization ............. 12 5.2.1 RO + Recirculation loop sanitization ......12 BASIC OPERATION ............16 USEFUL CONTACT DETAILS ........... 17 Page 1 PURELAB 7000 - US MANU40593 Version 02 - 02/16...
  • Page 4: Introduction

    1. INTRODUCTION Use of this Manual ELGA LabWater strives to produce manuals that are as simple and accurate as possible. However, should you feel they can be improved in any way please email us at info@elgalabwater.com. This manual guides you through the installation, commissioning and basic operation of the PURELAB 7000 allowing you to obtain a guaranteed supply of purified water to meet your requirements.
  • Page 5: Installation

    230 V, 50 Hz ac 13 amp 115 V, 60 Hz ac * Refer to Operator Manual for feed water specification. Suitability of water supply must be confirmed before commissioning is commenced. Page 3 PURELAB 7000 - US MANU40593 Version 02 - 02/16...
  • Page 6: Connecting Purelab 7000 - Us

    PURELAB 7000 – US Quick Start Manual ELGA Connecting PURELAB 7000 - US Once the PURELAB 7000 - US unit has been positioned, it should be connected as follows: CONNECT potable feed water supply using tube supplied. Notes:  Ensure a suitable valve is fitted to allow isolation of the supply during essential maintenance.
  • Page 7: Commissioning

     PRESS to shift cursor onto year.   PRESS to increase or to decrease year.  PRESS Set date screens PRESS ✔ . Page 5 PURELAB 7000 - US MANU40593 Version 02 - 02/16...
  • Page 8: Consumable Installation

    PURELAB 7000 – US Quick Start Manual ELGA 4. CONSUMABLE INSTALLATION Protek L1 or L2 Pre-treatment cartridge Screen prompt to install Protek pre-treatment cartridge (LC175 / LC177). OPEN front doors and locate the Protek L1 / L2 cartridge position as shown.
  • Page 9: Labpure L1 Purification Cartridge

    System purge PRESS the system now enters an automatic purge phase. After a maximum of 60 minutes the system will exit the purge phase regardless of the final water purity. Page 7 PURELAB 7000 - US MANU40593 Version 02 - 02/16...
  • Page 10: Sanitization

    PURELAB 7000 – US Quick Start Manual ELGA 5. SANITIZATION Liquid sanitization 5.1.1 RO + Recirculation loop sanitization Sanitization of the finished installation is essential to ensure that the system is properly commissioned and capable of achieving the bacteriological control required for a typical application.
  • Page 11 10. UNSCREW lid on the left of the unit. 11. Measure 50ml of Minncare cold sterilant and slowly pour into the break tank. 12. REPLACE lid. Large white lid removal Page 9 PURELAB 7000 - US MANU40593 Version 02 - 02/16...
  • Page 12 PURELAB 7000 – US Quick Start Manual ELGA 13. PRESENT GREEN sanitization key to confirm that the chemical have been added. Note: These two screens are shown alternately. PRESS left button to cancel. Confirmation chemical added Page 10 PURELAB 7000 – US MANU40593 Version 02 - 02/16...
  • Page 13 Note: The automated rinse cycles are designed to remove the chemical sanitant from the system but for critical applications a water test is advised prior to use to ensure user specific requirements are met. Set sanitization reminder Page 11 PURELAB 7000 - US MANU40593 Version 02 - 02/16...
  • Page 14: Tablet Sanitization

    PURELAB 7000 – US Quick Start Manual ELGA Tablet sanitization 5.2.1 RO + Recirculation loop sanitization Sanitization of the finished installation is essential to ensure that the system is properly commissioned and capable of achieving the bacteriological control required for a typical application.
  • Page 15 DOSAGE OF CHEMICAL IS USED. 10. UNSCREW lid on the left of the unit. Tablet cutter for cutting EfferSan™ tablet. 11. ADD ¼ of EfferSan™ tablets. 12. REPLACE lid. Large white lid removal Page 13 PURELAB 7000 - US MANU40593 Version 02 - 02/16...
  • Page 16 PURELAB 7000 – US Quick Start Manual ELGA 13. PRESENT GREEN sanitization key to confirm that the EfferSan™ tablets have been added. Note: These two screens are shown alternately. PRESS left button to cancel. Confirmation CT3 is added Page 14...
  • Page 17 Note: The automated rinse cycles are designed to remove the chemical sanitant from the system but for critical applications a water test is advised prior to use to ensure user specific requirements are met. Set sanitization reminder Page 15 PURELAB 7000 - US MANU40593 Version 02 - 02/16...
  • Page 18: Basic Operation

    Sanitization Procedure. PURELAB 7000 - US is supplied set up to operate on Monday to Friday, between the hours of 07:30 to 18:00. Refer to Operator Manual supplied on CD for instructions on how to adjust the times of operation or to switch to continuous recirculation.
  • Page 19: Useful Contact Details

    Fax: +44 (0) 203 567 7205 E-mail: techsupport@elgalabwater.com For the address of the nearest ELGA LabWater Sales and Service office visit the country list on our website. http//www.elgalabwater.com Or contact ELGA at the number above. ELGA is the global laboratory water brand name of Veolia Water.