CENTRA - S200 US/CENTRA - R200 US Operator Manual
Volume/flow measurement screen
Pressure measurement screen
Water purity screen
Uncompensated water quality
screen
Page 22
Step 5 - Volume/flow - displayed unit of measurement
Step 6 - Pressure - displayed unit of measurement
Step 7 - Water purity unit setting
Step 8 - Uncompensated water quality
CENTRA - S200 US/CENTRA - R200 US Version 01 – 08/14
The reservoir volume and recirculation flow can be
displayed in either liters (L) or US Gallons (USG).
1.
PRESS SCROLL
PRESS TICK  to proceed to step 6.
PRESS TICK .
2.
The loop pressure can be displayed in either bar (bar) or
pounds per square inch (psi).
1.
PRESS SCROLL
PRESS TICK  to proceed to step 7.
PRESS TICK .
2.
This screen allows preferred displayed units of water
purity to be set, to either, MΩ.cm or µS/cm. This only
affects the measurement in the recirculation loop.
1.
PRESS SCROLL
µS/cm)
PRESS TICK  to proceed to step 8.
PRESS TICK .
2.
A U will indicate uncompensated readings (recirculation
loop only) in the normal process screen.
1.
PRESS SCROLL
( = Uncompensated water quality ON)
PRESS TICK  to proceed to step 9.
PRESS TICK .
2.
Note:
The on-going display of uncompensated values is
generally not recommended as it can lead to
confusion
among
possibilities of using water of inadequate purity.
to change mode (L or USG)
OR
to change mode (bar or psi)
OR
to change mode (MΩ.cm or
OR
to change mode
OR
users
and
increase
ELGA
the