4.4.4 Alarm

It is a function to generate an alarm when the temperature is above or below the user-specified
temperature or when the door is opened [4.5.3 Refers to Alarm Display and Pop-up Notification
Function].
High Deviation
Temperature
Deviation
Temperature
Temperature
Door Open
Alarm Delay
52412Mechnicaloven_ver_1.0
Item
Set the upper limit of temperature
deviation.
If the actual temperature becomes
higher than the target temperature +
allowable upper limit temperature, an
alarm is generated.
Set the lower limit of temperature
deviation.
Low
If the actual temperature becomes
lower than the target temperature -
allowable lower limit temperature, an
alarm is generated.
Set the absolute upper limit
temperature.
High
An alarm occurs if the actual
temperature exceeds the absolute
upper limit temperature.
Set the alarm for the door open delay
time.
If the door opens exceeding Door
Open Alarm Delay, an alarm occurs.
Function
Input Value
Default
5
0 ~ 250℃
5
0 ~ 250℃
270
0 ~ 270℃
Off (0) ~ 30min
5