Please follow the guidelines listed below.
(1) Please install the instrument inside the laboratory with a little human access.
(2) Install it horizontally on a solid and flat floor.
(3) Install it in a place without any direct sunlight and heat sources.
(4) Please install it in a place without water or steam effect.
(5) It is recommended to install it in a place where there is an exhaust system such as a fume hood or
an arm hood.
(6) Please keep a certain space from the wall or other adjacent experimental equipment.
In particular, the door of the device opens in the left direction, please make sure to have enough space
on the left so that the door can be fully opened.
52412Mechnicaloven_ver_1.0
• Please install the instrument on a flat and solid floor for any case of an accident.
• Since the door is opened to the left with maximum 115°, please consider
sufficient installation space.
• Please keep the appropriate space (typically at least 30 cm) between other
devices or walls.