Cole Parmer OVF-800 Series User Manual page 41

Forced convection oven
Hide thumbs Also See for OVF-800 Series:
Table of Contents

Advertisement

Wait ON Timer is OFF / Wait OFF Timer is OFF
- The initial setting is shown as above.
- Timer set does not apply to this instrument.
Wait ON Timer is ON / Wait OFF Timer is OFF
- If you only set the Wait ON Timer as ON, the
instrument waits for the input value before raising
the temperature as the set value.
- In this case, the heater will be continuously
working to maintain the set value until the user
stops the instrument.
52412Mechnicaloven_ver_1.0
How to reserve the Wait ON/OFF timer
In case of
In case of
In case of
Wait ON Timer is OFF / Wait OFF Timer is ON
- If you only set the Wait OFF timer, the
temperature will maintain only for the entered
time once it reaches the target temperature.
- The tempreature control will be terminated if all
the reserved end time is passed.
In case of
Wait ON Timer is ON / Wait OFF Timer is ON
- If you turn on both Wait ON Timer and Wait OFF
Timer, you can set the delay time before starting
the heater and also set the maintaining time to
last the set temperature.
- In this case, the heater operation starts after
all the time set in the Wait ON has elapsed. After
reaching the target temperature set by the user,
the target temperature is maintained for the time
set by the Wait OFF. Temperature control is
terminated when all of the set Wait OFF time has
elapsed.
41

Advertisement

Table of Contents
loading

Table of Contents