Cole Parmer OVF-800 Series User Manual page 52

Forced convection oven
Hide thumbs Also See for OVF-800 Series:
Table of Contents

Advertisement

to relocation) with respect to the user's main operating temperature, Auto Tuning can be done for
accurate temperature control.
Auto Tuning temperature is performed by setting the target temperature value in the Basic Setting
Mode and starting the control, then activate the Auto Tuning button to ON in System Setup> Auto
Tuning.
[How to set Auto Tuning]
(1) After setting the temperature in Basic Mode, press the start button.
(2) It is executed when the ON button is activated in the System Settings> Auto Tuning screen. The
Auto Tuning function cannot be used when operation is stopped or in program control.
(3) It moves directly to the operation display window of Basic Mode, and the icon
on the top status bar blinks indicating that Auto Tuning is running.
(4) When Auto Tuning is completed, the icon
Tuning value is automatically saved.
Auto Tuning may take tens of minutes or hours depending on the set temperature.
Do not turn off the power of the device while Auto Tuning is in progress.
During Auto Tuning, it is recommended that the difference from the ambient temperature
be at least 30 degrees. If the Auto Tuning temperature is similar to the ambient
temperature, the reliability of the optimal parameter value may be reduced.
Auto Tuning is not available when temperature control is stopped in Basic Mode or under
temperature control in Program Mode.
52412Mechnicaloven_ver_1.0
on the top status bar disappears and the Auto
52

Advertisement

Table of Contents
loading

Table of Contents